The influence of electrode type, interelectrode distance (IED) and electrode orientation on EMG signals from the paraspinal muscles was investigated. Bipolar electrodes were placed at distances 2, 3, 4, 6 and 8 cm over the erector spinae in the cranio-caudal direction ("in series") as well as in the direction perpendicular to it ("in parallel"). Ten subjects performed 5 s isometric contractions of the erector spinae at 20, 40, 60, 80 and 100% MVC by pulling upward on a handlebar attached to the floor. RMS EMG signals were analyzed for mean average amplitude (AA). Mean total power (TP) and mean median frequency (MF) of the raw EMG signal were determined using fast Fourier transform. In addition to graded loading, sustained fatiguing contractions were performed from which TP and MF were obtained. With increasing IED the AA and TP increased while MF decreased. Although a trend towards higher AA, TP and MF was found for electrodes "in series", as compared to those "in parallel", the difference never reached significance. It is concluded that consistent information about muscle activity was obtained with Miniature Biopotential Skin Electrodes and 14445C Hewlett-Packard electrodes independently from IED or orientation. Orientation "in parallel" prevented the electrodes from sliding during muscle contraction. The third tested type, electrodes developed in the Neuromuscular Research Center, Boston, proved extremely sensitive to movement.Mesh:
Substances:
Year: 1997 PMID: 9402434
Source DB: PubMed Journal: Electromyogr Clin Neurophysiol ISSN: 0301-150X