MR microimaging of the lung using volume projection encoding.

M D Shattuck1, S L Gewalt, G H Glover, L W Hedlund, G A Johnson.   

Abstract

Radial acquisition (RA) techniques have been extended to produce isotropic, three-dimensional images of lung in live laboratory animals at spatial resolution down to 0.013 mm3 with a signal-to-noise ratio of 30:1. The pulse sequence and reconstruction algorithm have been adapted to allow acquisition of image matrices of up to 256(3) in less than 15 min. Scan-synchronous ventilation has been incorporated to limit breathing motion artifacts. The imaging sequence permits randomizing and/or discarding selected views to minimize the consequences of breathing motion. The signal in lung parenchyma was measured as a function of flip angle (alpha) for different repetition times and found to follow the predictions for which there is an optimum excitation (Ernst) angle. A single T1 relaxation value of 780 +/- 54 ms fits all data from six guinea pigs at 2.0 T. This T1 value parameterizes the signal and allows for a priori optimization, such as calculation of the Ernst angle appropriate for lung imaging.

Entities:  

Mesh:

Year:  1997        PMID: 9402195     DOI: 10.1002/mrm.1910380613

Source DB:  PubMed          Journal:  Magn Reson Med        ISSN: 0740-3194            Impact factor:   4.668


  6 in total

1.  MR image reconstruction algorithms for sparse k-space data: a Java-based integration.

Authors:  R de Beer; A Coron; D Graveron-Demilly; R Lethmate; S Nastase; D van Ormondt; F T A W Wajer
Journal:  MAGMA       Date:  2002-11       Impact factor: 2.310

Review 2.  Small animal imaging with magnetic resonance microscopy.

Authors:  Bastiaan Driehuys; John Nouls; Alexandra Badea; Elizabeth Bucholz; Ketan Ghaghada; Alexandra Petiet; Laurence W Hedlund
Journal:  ILAR J       Date:  2008

3.  Magnetic resonance imaging of disease progression and resolution in a transgenic mouse model of pulmonary fibrosis.

Authors:  Zackary I Cleveland; Yu M Zhou; Teckla G Akinyi; R Scott Dunn; Cynthia R Davidson; Jinbang Guo; Jason C Woods; William D Hardie
Journal:  Am J Physiol Lung Cell Mol Physiol       Date:  2017-01-27       Impact factor: 5.464

4.  Quantitative analysis of hyperpolarized 3He ventilation changes in mice challenged with methacholine.

Authors:  Nilesh N Mistry; Abraham Thomas; S Sivaram Kaushik; G Allan Johnson; Bastiaan Driehuys
Journal:  Magn Reson Med       Date:  2010-03       Impact factor: 4.668

5.  A Constant-Volume Ventilator and Gas Recapture System for Hyperpolarized Gas MRI of Mouse and Rat Lungs.

Authors:  John Nouls; Manuel Fanarjian; Laurence Hedlund; Bastiaan Driehuys
Journal:  Concepts Magn Reson Part B Magn Reson Eng       Date:  2011-04       Impact factor: 1.176

6.  3He MRI in mouse models of asthma.

Authors:  Bastiaan Driehuys; Julia Walker; Jim Pollaro; Gary P Cofer; Nilesh Mistry; David Schwartz; G Allan Johnson
Journal:  Magn Reson Med       Date:  2007-11       Impact factor: 4.668

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.