Literature DB >> 9393902

The cement mantle in femoral impaction allografting. A comparison of three systems from four centres.

E L Masterson1, B A Masri, C P Duncan, A Rosenberg, M Cabanela, M Gross.   

Abstract

An analysis of the cement mantle obtained with the Exeter impaction allografting system at one centre showed that it was either deficient or absent in almost 47% of Gruen zones. We therefore examined the mantle obtained using this system at another hospital and compared the results with those from the CPT and Harris Precoat Systems at other centres. The surgical indications for the procedure and the patient details were broadly similar in all four hospitals. There was some variation in the frequency of use of cortical strut allografts, cerclage wires and wire mesh to supplement the impaction allograft. Analysis of the cement mantles showed that when uncertain Gruen zones were excluded, the incidence of zones with areas of absence or deficiency of the cement was 47% and 50%, respectively, for the two centres using the Exeter system, 21% for the CPT system and 18% for the Harris Precoat system. We measured the difference in size between the proximal allograft impactors and the definitive prosthesis for each system. The Exeter system impactors are shorter than the definitive prosthesis and taper sharply so that the cavity created is inadequate, especially distally. The CPT proximal impactors are considerably longer than the definitive prosthesis and are designed to give a mantle of approximately 2 mm medially and laterally and 1.5 mm anteriorly and posteriorly. The Harris Precoat proximal impactors allow for a mantle with a circumference of 0.75 mm in the smaller sizes and 1 mm in the larger. Many reports link the longevity of a cemented implant to the adequacy of the cement mantle. For this reason, femoral impaction systems require careful design to achieve a cement mantle which is uninterrupted in its length and adequate in its thickness. Our results suggest that some current systems require modification.

Entities:  

Mesh:

Substances:

Year:  1997        PMID: 9393902     DOI: 10.1302/0301-620x.79b6.7690

Source DB:  PubMed          Journal:  J Bone Joint Surg Br        ISSN: 0301-620X


  3 in total

1.  Influence of antibiotic pellets on pore size and shear stress resistance of impacted native and thermodisinfected cancellous bone: An in vitro femoral impaction bone grafting model.

Authors:  C Fölsch; J Bok; G A Krombach; M Rickert; C A Fonseca Ulloa; G A Ahmed; M Kampschulte; A Jahnke
Journal:  J Orthop       Date:  2020-09-19

2.  Evaluation of Clinical Results and Complications of Structural Allograft Reconstruction after Bone Tumor Surgery.

Authors:  Mohammad Gharedaghi; Mohammad Taghi Peivandi; Mehdi Mazloomi; Hasan Rahimi Shoorin; Mohammad Hasani; Parham Seyf; Fatemeh Khazaee
Journal:  Arch Bone Jt Surg       Date:  2016-06

3.  In vitro testing of femoral impaction grafting with porous titanium particles: a pilot study.

Authors:  René Aquarius; Luc Walschot; Pieter Buma; Berend Willem Schreurs; Nico Verdonschot
Journal:  Clin Orthop Relat Res       Date:  2009-01-13       Impact factor: 4.176