| Literature DB >> 9389279 |
Abstract
The Bair Hugger system is a new and highly effective active patient warming system which produces a layer of warm air between the patient and the warming system. We report an instance of marked softening and distortion of a polyvinyl chloride tracheal tube caused by this layer. We also present laboratory data indicating that this is a likely problem under routine theatre conditions, with suggestions for prevention.Entities:
Mesh:
Substances:
Year: 1997 PMID: 9389279 DOI: 10.1093/bja/79.4.543
Source DB: PubMed Journal: Br J Anaesth ISSN: 0007-0912 Impact factor: 9.166