Comparison of manual and automated cell counts in EDTA preserved synovial fluids. Storage has little influence on the results.

M Salinas1, J Rosas, J Iborra, H Manero, E Pascual.   

Abstract

OBJECTIVE: To determine the precision and agreement of synovial fluid (SF) cell counts done manually and with automated counters, and to determine the degree of variability of the counts in SF samples, kept in the tubes used for routine white blood cell (WBC) counts--which use liquid EDTA as anticoagulant--at 24 and 48 hours at 4 degrees C, and at room temperature.
METHODS: To determine precision, cell counts were repeated 10 times--both manually and by an automated counter--in a SF sample of low, medium, and high cellularity. The variances were calculated to determine the interobserver variation in two manual (M1,M2) and two automated cell counts (C1,C2). The agreement between a manual (M1) and automated counter (C1) results, was analysed by the Bland and Altman method and the difference against the mean of the two methods was plotted. Then, the mean difference between the two methods was estimated and the standard deviation of the difference. To determine the effects of storage, SF samples were kept in a refrigerator at 4 degrees C, and at room temperature; cell counts were done manually (M1) and automatically (C1) at 24 and 48 hours and the changes analysed by the Bland and Altman method. The variances were compared using an F test.
RESULTS: (1) Precision. With the manual technique, the coefficients of variation were 27.9%, 14%, and 10.7% when used for counting the SF with low (270), medium (6200), and high cellularities (25,000). With the automated technique the coefficients of variation were 20%, 3.4%, and 2.9% in the same SF samples. In the fluids of medium and high cellularity, the variances of the automated cell counts were significatively lower (F test, p < 0.002) than those of the manual counts. (2) Interobserver variation. The variance between C1 and C2 (25 SF) was significatively lower (F test, p < 0.002) than that of the manual counts (41 SF). (3) Agreement between the two techniques (100 SF). For cellularities above 2000 cells/mm3, the manual method gave results between +10% to -34% of the results obtained by the coulter. For cellularities below 2000 cells/mm3, manual cell counts were between +60 to -1280 cells/mm3 of those obtained by the automated counter. (4) Influence of storage. The coulter counts of SF samples preserved at 4 degrees C showed less variance (F test, p < 0.05) than the manual counts. The worst results were obtained in manual counts of SF samples kept at room temperature; these samples at 48 hours showed a variation between -47% to 42% of the initial results.
CONCLUSIONS: Automated cell count of the SF offers advantages: it gives higher precision and consumes less time. The stability of the samples preserved in the EDTA tubes used for routine WBC counts is of additional interest, because if delay cannot be avoided, the results of the WBC counts are still accurate at 24 and even at 48 hours, at least for clinical purposes.

Entities:  

Mesh:

Substances:

Year:  1997        PMID: 9389224      PMCID: PMC1752268          DOI: 10.1136/ard.56.10.622

Source DB:  PubMed          Journal:  Ann Rheum Dis        ISSN: 0003-4967            Impact factor:   19.103


  9 in total

1.  Dried cytospin preparations of synovial fluid are a stable material for long-time storage and delayed crystal analysis.

Authors:  Christoph Robier; Manfred Neubauer; Mariana Stettin; Raimund Lunzer; Franz Rainer
Journal:  Clin Rheumatol       Date:  2012-03-14       Impact factor: 2.980

2.  The added value of synovial fluid centrifugation for monosodium urate and calcium pyrophosphate crystal detection.

Authors:  D Boumans; M E Hettema; H E Vonkeman; R G Maatman; M A van de Laar
Journal:  Clin Rheumatol       Date:  2017-04-19       Impact factor: 2.980

3.  Delayed examination of synovial fluid by ordinary and polarised light microscopy to detect and identify crystals.

Authors:  J Gálvez; E Sáiz; L F Linares; A Climent; C Marras; M F Pina; P Castellón
Journal:  Ann Rheum Dis       Date:  2002-05       Impact factor: 19.103

4.  6-Shogaol reduced chronic inflammatory response in the knees of rats treated with complete Freund's adjuvant.

Authors:  Arkene Sa Levy; Oswald Simon; Janet Shelly; Michael Gardener
Journal:  BMC Pharmacol       Date:  2006-10-01

5.  Collection of cerebrospinal fluid into EDTA versus plain tubes does not affect the standard analysis in dogs.

Authors:  Bodil Cathrine Koch; Lea Ophelia Daniels; Line Tang Thomsen; Michelle Brønniche Møller Nielsen; Mette Berendt; Hanne Gredal
Journal:  Acta Vet Scand       Date:  2019-05-06       Impact factor: 1.695

6.  Validation of an automated cell counting method for cGMP manufacturing of human induced pluripotent stem cells.

Authors:  Paolo Manzini; Valeria Peli; Araceli Rivera-Ordaz; Silvia Budelli; Mario Barilani; Lorenza Lazzari
Journal:  Biotechnol Rep (Amst)       Date:  2022-02-07

7.  Comparison of Two Methods for the Determination of the Effects of Ionizing Radiation on Blood Cell Counts in Mice.

Authors:  Ana L Romero-Weaver; Ann R Kennedy
Journal:  Int J Biomed Sci       Date:  2012-03-15

Review 8.  Laboratory testing of extravascular body fluids: National recommendations on behalf of the Croatian Society of Medical Biochemistry and Laboratory Medicine. Part II - Synovial fluid.

Authors:  Anja Jokic; Lara Milevoj Kopcinovic; Jelena Culej; Irena Kocijan; Marija Bozovic
Journal:  Biochem Med (Zagreb)       Date:  2020-08-05       Impact factor: 2.313

9.  The Graphical Representation of Cell Count Representation: A New Procedure for the Diagnosis of Periprosthetic Joint Infections.

Authors:  Bernd Fink; Marius Hoyka; Elke Weissbarth; Philipp Schuster; Irina Berger
Journal:  Antibiotics (Basel)       Date:  2021-03-24
  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.