Literature DB >> 938441

[Recording of ventricular pressure by conventional catheter manometer systems. I. Minimal requirements of blood pressure recording systems and estimation of frequency response characteristics].

G Hellige.   

Abstract

Parallel recordings of pressure pulses by conventional catheter manometer systems and catheter tip manometer demonstrate severe errors in the peak velocity of pressure rise estimated by conventional systems. This fact is due to inadequate dynamic response characteristics of conventional systems in relation to the frequency content of pressure curves. During cardiac rest the error in dp/dt max is less than 10% if the frequency response of the recording system is uniform up to 10 Hz, the corresponding value under maximal cardiac stimulation is about 40 Hz. This is equal to the first 10 harmonics of heart rate. The examination of left and right ventricular pressure curves leads to similar results. The experimental determination of dynamic response characteristics of cathermanometer systems requires a test system producing suitable sinus or step functions, parallel high fidelity recording of pressure functions to be recommended. A simple test station is described. Examinations of temperature influence on catheter material and resulting changes in dynamic response characteristics were carried out. The incubation of catheters at the temperature of 37 degrees C is indispensable. A new diagram for simplifying the interpretation of results is described. Other publications are discussed in viewpoint of employed techniques and representation of results.

Entities:  

Mesh:

Year:  1976        PMID: 938441     DOI: 10.1007/bf01906457

Source DB:  PubMed          Journal:  Basic Res Cardiol        ISSN: 0300-8428            Impact factor:   17.165


  9 in total

1.  HARMONIC ANALYSIS OF PRESSURE PULSES OBTAINED FROM THE HEART AND GREAT VESSELS OF MAN.

Authors:  D J PATEL; D T MASON; J ROSS; E BRAUNWALD
Journal:  Am Heart J       Date:  1965-06       Impact factor: 4.749

2.  Special instrumentation problems encountered in physiological research concerning the heart and circulation in man.

Authors:  E H WOOD
Journal:  Science       Date:  1950-12-15       Impact factor: 47.728

3.  Thermodilution cardiac output determination with a single flow-directed catheter.

Authors:  J S Forrester; W Ganz; G Diamond; T McHugh; D W Chonette; H J Swan
Journal:  Am Heart J       Date:  1972-03       Impact factor: 4.749

4.  [Recording accuracy of various heart catheters. Comparative studies using a catheter tip manometer in the frequency range of 1 to 10 Hertz].

Authors:  F H Degenring; H A Dittmar
Journal:  Z Kardiol       Date:  1973-08

5.  [Early diagnosis of heart insufficiency using a floating catheter system. Studies on 15 myocardial infarct patients 3 to 6 weeks and 10 patients more than 4 months following the acute attack].

Authors:  H Roskamm; H Weidemann; B Meinecke; J Petersen; H Reindell
Journal:  Z Kreislaufforsch       Date:  1970-02

6.  Damped and undamped frequency responses of underdamped catheter manometer systems.

Authors:  G G Shapiro; L J Krovetz
Journal:  Am Heart J       Date:  1970-08       Impact factor: 4.749

7.  Evaluation of the contractile state of the human heart.

Authors:  D T Mason; R Zelis; E A Amsterdam
Journal:  Triangle       Date:  1970

8.  [Comparative metrological studies of the Cournand heart catheter and microcatheter].

Authors:  W Weller; W T Ulmer; W Walkenhorst
Journal:  Z Kreislaufforsch       Date:  1969-07

9.  A simple, inexpensive, sinusoidal pressure generator.

Authors:  H F Stegall
Journal:  J Appl Physiol       Date:  1967-03       Impact factor: 3.531

  9 in total
  2 in total

1.  A critical look at currently used indirect indices of myocardial oxygen consumption.

Authors:  D Baller; H J Bretschneider; G Hellige
Journal:  Basic Res Cardiol       Date:  1981 Mar-Apr       Impact factor: 17.165

2.  Availability of intraarterial pressure waveforms from catheter-manometer systems during surgery.

Authors:  K H Wesseling; N T Smith
Journal:  J Clin Monit       Date:  1985-01
  2 in total