P S Hodder1, G Blankenstein, J Ruzicka.
Abstract
A microfabricated flow chamber (MFC) suitable for performing liquid-based fluorimetric assays is introduced. Precision delivery of microliter volumes of sample and reagent to the MFC is accomplished by a double-syringe-pump flow injection analysis (FIA) apparatus. The FIA-MFC system also combines the 'sheath flow' technique (traditionally used in flow cytometry) and stopped-flow FIA as a way to allow sample and reagent streams to be mixed reproducibly. The applicability of this FIA-MFC system to bioanalytical assays is demonstrated by performing an enzymatic assay with an artificial fluorigenic substrate to determine the activity of Savinase, a proteolytic enzyme. When coupled to a fluorescence microscope platform, quantitative analysis of the reaction product is possible. Experiments showed that the FIA-MFC system was capable of performing the assay with good reproducibility of injection (1.5%), and linearity of response (r2 = 0.9997) in activity ranges of analytical interest. Owing to the incorporation of flow cytometry sheath flow principles into an FIA format, the FIA-MFC system is a suitable tool for cytometric studies.Mesh:
Substances:
Year: 1997 PMID: 9374023 DOI: 10.1039/a701750b
Source DB: PubMed Journal: Analyst ISSN: 0003-2654 Impact factor: 4.616