Literature DB >> 9338637

Clinical and laboratory evaluation of a new thin-walled self-expanding tracheobronchial silicone stent: progress and pitfalls.

K Wassermann1, A Koch, J Müller-Ehmsen, M Reuter, O Michel, H E Eckel.   

Abstract

BACKGROUND: Although widely established in the management of malignant airway lesions, currently available tracheobronchial prostheses made of silicone have their drawbacks resulting from rigidity and wall thickness. Therefore we present clinical follow-up data obtained with a novel thin-walled expandable prototype silicone airway stent (Polyflex stent, Willy Rüsch AG, Kernen, Germany) in 19 patients.
METHODS: Seventeen of 19 patients had tracheobronchial complications of infiltrating cancer: five had respiratory-digestive fistulas, 14 had mixed-type obstructions (mucosal infiltration plus extrinsic compression), and two had diffuse tracheal hemorrhages from the tumor surface (three patients had more than one complication). Two of 19 patients had benign postintubation stricture and malacia. Overall, 33 stents were implanted either simultaneously or in a consecutive manner. Scanning electron microscopy was performed both on prototype stents and on other available silicone stents for comparison.
RESULTS: The treatment improved the patients' clinical condition substantially. The mechanical properties of the new prosthesis were excellent. Important stent-associated side effects were early mucus retention (n = 7), infolding of the inner silicone layer (n = 2), and stent dislodgment (n = 2). As of February 1997, 10 patients have died of causes unrelated to stent placement. Seven patients with malignant airway disease are still alive from 2 weeks up to 7 months after initial treatment. Scanning electron microscopy of explanted and unused prototypes suggested that an extremely ragged luminal microstructure may contribute to the firm adhesion of secretory material and that technical smoothing of the surface avoids such complications.
CONCLUSIONS: The novel self-expandable silicone airway stent may be a promising addition to commonly used stent types. Short-term and medium-term management of fistulas, tumor surface bleeding, and strictures (malignant and benign) is satisfactory. Scanning electron microscopy of stents provides information on peculiar features of microstructure and material that may be of use in clinical research and technical innovation.

Entities:  

Mesh:

Substances:

Year:  1997        PMID: 9338637     DOI: 10.1016/S0022-5223(97)70040-0

Source DB:  PubMed          Journal:  J Thorac Cardiovasc Surg        ISSN: 0022-5223            Impact factor:   5.209


  4 in total

1.  [Clinical evaluation of Dumon stent for tracheobronchial stenosis due to advanced esophageal carcinoma].

Authors:  M Ohara; T Abe; Y Tanabe; M Kusano; T Kusumi; T Okayasu; M Hosokawa
Journal:  Jpn J Thorac Cardiovasc Surg       Date:  1998-11

2.  Progressive experience in tracheal stenting with self-expandable stents.

Authors:  Marc Remacle; Georges Lawson; Jacques Jamart; Jerôme Keghian
Journal:  Eur Arch Otorhinolaryngol       Date:  2003-03-18       Impact factor: 2.503

3.  Expandable polyester silicon-covered stent for malignant esophageal strictures before neoadjuvant chemoradiation: a pilot study.

Authors:  Ali A Siddiqui; David Loren; Robert Dudnick; Thomas Kowalski
Journal:  Dig Dis Sci       Date:  2007-03       Impact factor: 3.487

4.  A Safe Method of Tracheal Polyflex Stent Placement: A Review of 20 Patients.

Authors:  Saviz Pejhan; Mojtaba Javaherzadeh; Abolghasem Daneshvar; Roya Farzanegan
Journal:  Iran Red Crescent Med J       Date:  2015-08-29       Impact factor: 0.611

  4 in total