| Literature DB >> 9330544 |
R Hauser1, B Westermann, R Probst.
Abstract
A basic problem common to all systems for computer assisted surgery (CAS) is patient referencing, or the transfer of preoperative image data to the intraoperative pathology. The authors describe a highly precise CAS system with non-invasive referencing that can be used in ear, nose and throat (ENT) surgery of the paranasal sinuses. It is based on optical digitizing with several custom-made self-localizing surgical instruments. The accuracy of the system was tested in an experimental model using a plastic head. Measurements of repositioning the reference bow had a mean error of 0.81 mm +/- 0.31 mm. The system was evaluated clinically with 11 patients who received surgery for different pathologies of the paranasal sinuses. These trials met with a high rate of success and specific results are reported.Entities:
Mesh:
Year: 1997 PMID: 9330544 DOI: 10.1243/0954411971534458
Source DB: PubMed Journal: Proc Inst Mech Eng H ISSN: 0954-4119 Impact factor: 1.617