Literature DB >> 9330181

Biocompatibility of silver-modified polyester for antimicrobial protection of prosthetic valves.

K S Tweden1, J D Cameron, A J Razzouk, W R Holmberg, S J Kelly.   

Abstract

BACKGROUND AND AIMS OF THE STUDY: The biocompatibility of a silver-coated polyethylene terephthlate (PET, polyester) fabric for the inhibition of prosthetic valve endocarditis (PVE) associated with mechanical heart valves (MHVs) was assessed. The infrequency of PVE is outweighed by mortality rates commonly exceeding 50%. These high mortality rates have been attributed to the poor effect of antibiotic therapy on colonized valves and infected myocardial tissue. Silver has been used as an antimicrobial for centuries due to its general effectiveness and relative lack of toxicity. Our previous work has shown PET polyester fabric coated with metallic silver by an ion beam-assisted deposition (IBAD) process to: (i) be effective in vitro in the inhibition of microbial attachment and colonization; (ii) be tightly adherent and low leaching; and (iii) promote tissue ingrowth and the organization of tissue pannus in a short-duration (five weeks) sheep mitral mechanical heart valve model.
METHODS: This paper addresses additional biocompatibility assessment consisting of a cell compatibility assay in which serum extracts of silver-coated fabric were exposed to fibroblasts for 48 hours, after which cell viability and function were measured. The amount of silver in the extract was measured using elemental analysis techniques.
RESULTS: No signs of toxicity were seen in the cells until the extract concentration reached 1200 p.p.m. Ten-week duration mechanical valve replacement studies in sheep with uncoated or coated polyester sewing cuffs showed comparable tissue ingrowth and mature pannus with a suggestion of a thinner pannus on the silver-coated fabric. Additional antimicrobial testing confirmed the effectiveness of this coating in inhibiting colonization of polyester fabric.
CONCLUSIONS: These current results, together with the earlier data, suggest that IBAD silver coating on polyester facilitates healing and may provide protection against PVE.

Entities:  

Mesh:

Substances:

Year:  1997        PMID: 9330181

Source DB:  PubMed          Journal:  J Heart Valve Dis        ISSN: 0966-8519


  15 in total

Review 1.  Silver-coated megaprostheses: review of the literature.

Authors:  Tom Schmidt-Braekling; Arne Streitbuerger; Georg Gosheger; Friedrich Boettner; Markus Nottrott; Helmut Ahrens; Ralf Dieckmann; Wiebke Guder; Dimosthenis Andreou; Gregor Hauschild; Burkhard Moellenbeck; Wenzel Waldstein; Jendrik Hardes
Journal:  Eur J Orthop Surg Traumatol       Date:  2017-03-06

Review 2.  Silver-coated megaprosthesis in prevention and treatment of peri-prosthetic infections: a systematic review and meta-analysis about efficacy and toxicity in primary and revision surgery.

Authors:  Michele Fiore; Andrea Sambri; Riccardo Zucchini; Claudio Giannini; Davide Maria Donati; Massimiliano De Paolis
Journal:  Eur J Orthop Surg Traumatol       Date:  2020-09-05

3.  [Nanoparticulate silver. A new antimicrobial substance for bone cement].

Authors:  V Alt; T Bechert; P Steinrücke; M Wagener; P Seidel; E Dingeldein; D Scheddin; E Domann; R Schnettler
Journal:  Orthopade       Date:  2004-08       Impact factor: 1.087

4.  Incidence of embolism and paravalvar leak after St Jude Silzone valve implantation: experience from the Cardiff Embolic Risk Factor Study.

Authors:  A Ionescu; N Payne; A G Fraser; J Giddings; G L Grunkemeier; E G Butchart
Journal:  Heart       Date:  2003-09       Impact factor: 5.994

5.  Evaluation of osseous integration of PVD-silver-coated hip prostheses in a canine model.

Authors:  Gregor Hauschild; Jendrik Hardes; Georg Gosheger; Sandra Stoeppeler; Helmut Ahrens; Franziska Blaske; Christoph Wehe; Uwe Karst; Steffen Höll
Journal:  Biomed Res Int       Date:  2015-01-28       Impact factor: 3.411

6.  Measurement of the silver ion concentration in wound fluids after implantation of silver-coated megaprostheses: correlation with the clinical outcome.

Authors:  B Hussmann; I Johann; M D Kauther; S Landgraeber; Marcus Jäger; S Lendemans
Journal:  Biomed Res Int       Date:  2013-05-29       Impact factor: 3.411

7.  The influence of elementary silver versus titanium on osteoblasts behaviour in vitro using human osteosarcoma cell lines.

Authors:  Jendrik Hardes; Arne Streitburger; Helmut Ahrens; Thomas Nusselt; Carsten Gebert; Winfried Winkelmann; Achim Battmann; Georg Gosheger
Journal:  Sarcoma       Date:  2007

8.  Anticancer Activity of Ag(I) N-Heterocyclic Carbene Complexes Derived from 4,5-Dichloro-1H-Imidazole.

Authors:  Doug A Medvetz; Khadijah M Hindi; Matthew J Panzner; Andrew J Ditto; Yang H Yun; Wiley J Youngs
Journal:  Met Based Drugs       Date:  2008

9.  Bacteriophage mediated killing of Staphylococcus aureus in vitro on orthopaedic K wires in presence of linezolid prevents implant colonization.

Authors:  Sandeep Kaur; Kusum Harjai; Sanjay Chhibber
Journal:  PLoS One       Date:  2014-03-03       Impact factor: 3.240

10.  A new model of implant-related osteomyelitis in the metaphysis of rat tibiae.

Authors:  Norbert Harrasser; Johannes Gorkotte; Andreas Obermeier; Susanne Feihl; Melanie Straub; Julia Slotta-Huspenina; Ruediger von Eisenhart-Rothe; Walter Moser; Philipp Gruner; Michael de Wild; Hans Gollwitzer; Rainer Burgkart
Journal:  BMC Musculoskelet Disord       Date:  2016-04-08       Impact factor: 2.362

View more