Multicenter study on the clinical value of fetal pulse oximetry. II. Compared predictive values of pulse oximetry and fetal blood analysis. The French Study Group on Fetal Pulse Oximetry.

B Carbonne1, B Langer, F Goffinet, F Audibert, D Tardif, F Le Goueff, M Laville, F Maillard.   

Abstract

OBJECTIVE: Our purpose was to compare the predictive value of intrapartum fetal pulse oximetry with that of fetal blood analysis for an abnormal neonatal outcome in case of an abnormal fetal heart rate. STUDY
DESIGN: A prospective multicenter observational study was conducted from June 1994 to November 1995. Fetal oxygen saturation was continuously recorded with a Nellcor N-400 fetal pulse oximeter in case of an abnormal fetal heart rate during labor. Simultaneous readings of fetal oxygen saturation and fetal blood analysis obtained before birth (i.e., either at full dilatation or before cesarean section when indicated) were compared with the neonatal status. The criteria for an abnormal neonatal outcome were (1) an umbilical arterial blood pH < or = 7.15 and (2) a combined variable including 5-minute Apgar score < or = 7, umbilical arterial pH < or = 7.15, secondary respiratory distress, transfer in a neonatal care unit, or neonatal death.
RESULTS: At a 7.20 threshold for fetal scalp pH and 30% for fetal oxygen saturation (i.e., the 10th percentile in the study population), the predictive value of fetal pulse oximetry was similar to that of fetal blood analysis for an arterial umbilical pH < or = 7.15 and for an abnormal neonatal outcome (positive predictive value 56% vs 55%, negative predictive value 81% vs 82%, sensitivity 29% vs 35%, and specificity 93% vs 91%, respectively). The receiver-operator characteristic curve showed similar performance of either technique for cutoff values < or = 7.20 for fetal blood pH and < or = 30% for fetal oxygen saturation, whereas fetal pulse oximetry became superior at higher thresholds.
CONCLUSION: The predictive value of intrapartum fetal pulse oximetry can be favorably compared with that of fetal blood analysis. Randomized controlled management trials can now be performed to assess potential clinical benefits of this new tool.

Entities:  

Mesh:

Year:  1997        PMID: 9322629     DOI: 10.1016/s0002-9378(97)70151-0

Source DB:  PubMed          Journal:  Am J Obstet Gynecol        ISSN: 0002-9378            Impact factor:   8.661


  2 in total

Review 1.  Continuous cardiotocography (CTG) as a form of electronic fetal monitoring (EFM) for fetal assessment during labour.

Authors:  Zarko Alfirevic; Declan Devane; Gillian Ml Gyte; Anna Cuthbert
Journal:  Cochrane Database Syst Rev       Date:  2017-02-03

2.  Validation of noninvasive photoacoustic measurements of sagittal sinus oxyhemoglobin saturation in hypoxic neonatal piglets.

Authors:  Jeeun Kang; Emad M Boctor; Shawn Adams; Ewa Kulikowicz; Haichong K Zhang; Raymond C Koehler; Ernest M Graham
Journal:  J Appl Physiol (1985)       Date:  2018-06-21