A stress analysis of the femoral stem in total hip prostheses.

T P Andriacchi, J O Galante, T B Belytschko, S Hampton.   

Abstract

A two-dimensional stress analysis was used to study the effects of some of the factors leading to early fatigue failure of the femoral stem in total hip prostheses. The results show that loss of proximal stem support at the level of the calcar femorale will result in stem stress levels which can lead to fatigue failure. In addition, the role of body weight and range of cyclic stress fluctuation play an important role in fatigue life under conditions where the stem has lost proximal support. These results indicate that stem design could be improved by incorporating some means of ensuring adequate support at the calcar femoralre maximum tensile stresses are found to occur.

Entities:  

Mesh:

Year:  1976        PMID: 932061

Source DB:  PubMed          Journal:  J Bone Joint Surg Am        ISSN: 0021-9355            Impact factor:   5.284


  13 in total

1.  Early failure of a non-cemented femoral stem after minimal-invasive total hip arthroplasty: cause analysis and classification.

Authors:  Jan P Benthien; Wolf Siepen
Journal:  Musculoskelet Surg       Date:  2012-05-06

2.  Laser etching causing fatigue fracture at the neck-shoulder junction of an uncemented femoral stem: A case report.

Authors:  Bob Jang; Andrew Kanawati; Declan Brazil; Warwick Bruce
Journal:  J Orthop       Date:  2013-05-24

3.  A physico-mathematical model for the human femur, with and without a prosthesis, under the static constraints of one-legged stance.

Authors:  J C Ferré; R Legoux; F Marquet; C Chevalier; J L Helary; J P Lumineau; A Y Le Cloarec; E Orio; J G Barbin; J Y Barbin
Journal:  Surg Radiol Anat       Date:  1987       Impact factor: 1.246

4.  The effects of hip contact aberrations on stress patterns within the human femoral head.

Authors:  T D Brown; A B Ferguson
Journal:  Ann Biomed Eng       Date:  1980       Impact factor: 3.934

5.  Rare fracture of a total hip replacement.

Authors:  G Aldinger; K Mitzkat
Journal:  Arch Orthop Trauma Surg       Date:  1983

6.  Importance of a distal centralizer in experimental malpositioning of cemented stems. A biomechanical study on human femora.

Authors:  Andreas Kusserow; Andreas Ficklscherer; Peter Cornelius Kreuz; Susanne Finze; Wolfram Mittelmeier; Volkmar Jansson; Stefan Milz; Bernd Wegener
Journal:  Arch Med Sci       Date:  2015-12-11       Impact factor: 3.318

7.  Fatigue fracture of a cemented Omnifit CoCr femoral stem: implant and failure analysis.

Authors:  Noah Bonnheim; Hannah Gramling; Michael Ries; Sanjai Shukla; Beatrice Iliescu; Lisa Pruitt
Journal:  Arthroplast Today       Date:  2017-08-02

8.  Selection of an animal model for implant fixation studies: anatomical aspects.

Authors:  V K Goel; H Drinker; M M Panjabi; A Strongwater
Journal:  Yale J Biol Med       Date:  1982 Mar-Apr

9.  A simple new technique for the removal of fractured femoral stems: a case report.

Authors:  Goran Bicanic; Kresimir Crnogaca; Domagoj Delimar
Journal:  J Med Case Rep       Date:  2014-05-15

10.  Fracture of Uncemented Revision Femoral Stems in three Arthroplasty Patients: A Case Series with three different brands.

Authors:  Rajiv Merchant; Ian Kelly; John Quinlan
Journal:  J Orthop Case Rep       Date:  2014 Oct-Dec
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.