[Fully automated generation of hip prosthesis voxel models of the femur for finite element analysis. Comparison of direct and subsequent prosthesis implantation].

J Schmitt1, M Lengsfeld, R Leppek, P Alter.   

Abstract

Two means of generating implant-bearing FE voxel models of the human femur are described and compared. The first method consists in the fully automatic production of a femur model from CT data and subsequent implantation of a prosthesis stem on the basis of a redefinition of existing voxel elements using geometrical data control. The second method processes the CT data of a femur already bearing a prosthesis. Here, the prosthesis elements are generated separately from the surrounding bony elements on the basis of the differing CT density thresholds. Although the two implantation algorithms produce comparable results, they have different fields of application. The method of numerical stress analysis involving automatic generation of voxel models can be applied to implant-bearing femurs. Clinical follow-up observations of patients with prostheses made of titanium can be supported by FE calculations.

Entities:  

Mesh:

Year:  1997        PMID: 9312305     DOI: 10.1515/bmte.1997.42.6.150