Distribution of inhaled nitric oxide during sequential and continuous administration into the inspiratory limb of the ventilator.

E Mourgeon1, L Gallart, G S Rao, Q Lu, J D Law-Koune, L Puybasset, P Coriat, J J Rouby.   

Abstract

OBJECTIVES: The concentrations of nitric oxide (NO) in the ventilatory circuits and the patient's airways were compared between sequential (SQA) and continuous (CTA) administration during inspiratory limb delivery.
DESIGN: Prospective controlled study.
SETTING: 14-bed Surgical Intensive Care Unit of a teaching University hospital. PATIENTS AND PARTICIPANTS: Eleven patients with acute lung injury on mechanical ventilation and two healthy volunteers.
INTERVENTIONS: A prototype NO delivery device (Opti-NO) and César ventilator were set up in order to deliver 1, 3 and 6 parts per million (ppm) of NO into the bellows of a lung model in SQA and CTA. Using identical ventilatory and Opti-NO settings, NO was administered to the patients with acute lung injury. MEASUREMENTS AND
RESULTS: NO concentrations measured from the inspiratory limb [INSP-NOMeas] and the trachea [TRACH-NOMeas] using fast response chemiluminescence were compared between the lung model and the patients using controlled mechanical ventilation with a constant inspiratory flow. INSP-NOMeas were stable during SQA and fluctuated widely during CTA (fluctuation at 6 ppm = 61% in the lung model and 58 +/- 3% in patients). In patients, [TRACH-NOMeas] fluctuated widely during both modes (fluctuation at 6 ppm = 55 +/- 3% during SQA and 54 +/- 5% during CTA). The NO flow requirement was significantly lower during SQA than during CTA (74 +/- 0.5 vs 158 +/- 2.2 ml.min-1 to attain 6 ppm, p = 0.0001). INSP-NOMeas were close to the values predicted using a classical formula only during SQA (bias = -0.1 ppm, precision = +/-1 ppm during SQA; bias = 2.93 ppm and precision = +/-3.54 ppm during CTA). During SQA, INSP-NOMeas varied widely in healthy volunteers on pressure support ventilation.
CONCLUSIONS: CTA did not provide homogenous mixing of NO with the tidal volume and resulted in fluctuating INSP-NOMeas. In contrast, SQA delivered stable and predictable NO concentrations during controlled mechanical ventilation with a constant inspiratory flow and was economical compared to CTA. However, SQA did not provide stable and predictable NO concentrations during pressure support ventilation.

Entities:  

Mesh:

Substances:

Year:  1997        PMID: 9310802     DOI: 10.1007/s001340050421

Source DB:  PubMed          Journal:  Intensive Care Med        ISSN: 0342-4642            Impact factor:   17.440


  6 in total

1.  Pulmonary uptake and modes of administration of inhaled nitric oxide in mechanically-ventilated patients.

Authors: 
Journal:  Crit Care       Date:  1998       Impact factor: 9.097

2.  Inhaled nitric oxide fraction is influenced by both the site and the mode of administration.

Authors:  E Sieffert; L Ducros; M R Losser; D M Payen
Journal:  J Clin Monit Comput       Date:  1999-12       Impact factor: 2.502

3.  Volatile Anesthetics and the Treatment of Severe Bronchospasm: A Concept of Targeted Delivery.

Authors:  Jarred R Mondoñedo; John S McNeil; Samir D Amin; Jacob Herrmann; Brett A Simon; David W Kaczka
Journal:  Drug Discov Today Dis Models       Date:  2014-04-29

4.  Inhaled nitric oxide in acute respiratory distress syndrome with and without septic shock requiring norepinephrine administration: a dose-response study.

Authors: 
Journal:  Crit Care       Date:  1997       Impact factor: 9.097

5.  Fluctuations of inspired concentrations of nitric oxide and nitrogen dioxide during mechanical ventilation.

Authors: 
Journal:  Crit Care       Date:  1999       Impact factor: 9.097

6.  An injection and mixing element for delivery and monitoring of inhaled nitric oxide.

Authors:  Andrew R Martin; Chris Jackson; Samuel Fromont; Chloe Pont; Ira M Katz; Georges Caillobotte
Journal:  Biomed Eng Online       Date:  2016-08-30       Impact factor: 2.819

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.