Prophylactic balloon angioplasty fails to prolong the patency of expanded polytetrafluoroethylene arteriovenous grafts: results of a prospective randomized study.

A B Lumsden1, M J MacDonald, D Kikeri, G A Cotsonis, L A Harker, L G Martin.   

Abstract

PURPOSE: Maintenance of hemodialysis access grafts represents an enormous social and clinical problem. Current grafts and graft salvage techniques are inadequate. Consequently, there has been increasing interest in the use of minimally invasive catheter techniques to prophylactically treat stenoses in functioning arteriovenous grafts. Prophylactic balloon angioplasty has been widely suggested as prolonging assisted primary patency. We have performed a prospective randomized trial to compare patients who underwent percutaneous transluminal angioplasty (PTA) for graft stenoses > 50% with a control group that received no intervention. Our hypothesis was that to be efficacious a minimal benefit of 20% prolongation in patency would be necessary.
METHODS: Color flow duplex scanning was used to detect > 50% stenoses in functioning expanded polytetrafluoroethylene grafts. Patients were then subjected to confirmatory angiographic evaluation. Those who had angiographic stenoses > 50% were randomized to balloon angioplasty or observation. Patients were followed-up with duplex scanning every 2 months. Statistical analysis was performed using the Kaplan-Meier technique. Although demographically the patient groups were well matched, there were more prior interventions and concurrent central stenoses in the treatment group. Outcomes were graft thrombosis, graft dysfunction that precluded dialysis, and six or more PTA procedures within 18 months.
RESULTS: In the treatment and observation groups, the 6-month patency rates were 69% +/- 7% and 70% +/- 7%, respectively. The 12-month patency rates for the treatment and observation groups were 51% +/- 6% and 47% +/- 4%, respectively. There was no significant difference between these two groups (p = 0.97), with an 80% confidence limit for detection of a difference greater than 20%.
CONCLUSIONS: This study demonstrates that a generic approach of PTA to treat all polytetrafluoroethylene grafts with stenoses > 50% does not prolong patency and cannot be supported.

Entities:  

Mesh:

Substances:

Year:  1997        PMID: 9308584     DOI: 10.1016/s0741-5214(97)70031-4

Source DB:  PubMed          Journal:  J Vasc Surg        ISSN: 0741-5214            Impact factor:   4.268


  11 in total

Review 1.  Stent and Stent-Graft Use in Arteriovenous Dialysis Access.

Authors:  Gordon McLennan
Journal:  Semin Intervent Radiol       Date:  2016-03       Impact factor: 1.513

2.  A Patient with Recurrent Arteriovenous Graft Thrombosis.

Authors:  Michael Allon
Journal:  Clin J Am Soc Nephrol       Date:  2015-04-16       Impact factor: 8.237

3.  Surveillance of hemodialysis vascular access.

Authors:  William L Whittier
Journal:  Semin Intervent Radiol       Date:  2009-06       Impact factor: 1.513

Review 4.  Vascular access for hemodialysis: postoperative evaluation and function monitoring.

Authors:  Konstantinos Leivaditis; Stelios Panagoutsos; Athanasios Roumeliotis; Vassilios Liakopoulos; Vassilis Vargemezis
Journal:  Int Urol Nephrol       Date:  2013-09-18       Impact factor: 2.370

Review 5.  Novel therapies for hemodialysis vascular access dysfunction: myth or reality?

Authors:  Christi M Terry; Laura M Dember
Journal:  Clin J Am Soc Nephrol       Date:  2013-11-14       Impact factor: 8.237

Review 6.  Pre-emptive correction for haemodialysis arteriovenous access stenosis.

Authors:  Pietro Ravani; Robert R Quinn; Matthew J Oliver; Divya J Karsanji; Matthew T James; Jennifer M MacRae; Suetonia C Palmer; Giovanni F M Strippoli
Journal:  Cochrane Database Syst Rev       Date:  2016-01-07

Review 7.  Hemodialysis vascular access monitoring: current concepts.

Authors:  Michael Allon; Michelle L Robbin
Journal:  Hemodial Int       Date:  2009-04       Impact factor: 1.812

8.  Elbow fistulas using autogeneous vein: patency rates and results of revision.

Authors:  G J Murphy; R Saunders; M Metcalfe; M L Nicholson
Journal:  Postgrad Med J       Date:  2002-08       Impact factor: 2.401

9.  Surveillance and monitoring of dialysis access.

Authors:  Lalathaksha Kumbar; Jariatul Karim; Anatole Besarab
Journal:  Int J Nephrol       Date:  2011-11-22

10.  Effect of online haemodialysis vascular access flow evaluation and pre-emptive intervention on the frequency of access thrombosis.

Authors:  Edwin Wijnen; Frank M van der Sande; Jan H M Tordoir; Jeroen P Kooman; Karel M L Leunissen
Journal:  NDT Plus       Date:  2008-08-22
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.