Literature DB >> 9307952

Practical implementation of enhanced dynamic wedge in the CadPlan treatment planning system.

A Samuelsson1, K A Johansson, O Mattsson, A Palm, H Puurunen, G Sernbo.   

Abstract

The Varian CadPlan algorithm for computation of relative dose distributions and monitor unit calculations for Enhanced Dynamic Wedge (EDW) fields is based on a combination of open field beam data and Segmented Treatment data Tables. Calculation of dose by the pencil beam convolution model uses scatter kernels and boundary kernels to create the distribution. The principles of the pencil beam convolution model is presented. Comparison of measured and calculated monitor units and relative dose distributions showed good agreement and the deviations are within international accepted tolerans. Test results indicate that the EDW model works satisfactorily for all energies and wedge angles.

Mesh:

Year:  1997        PMID: 9307952     DOI: 10.1016/s0958-3947(97)00019-8

Source DB:  PubMed          Journal:  Med Dosim        ISSN: 1873-4022            Impact factor:   1.482


  1 in total

1.  Comparison between measured and calculated dynamic wedge dose distributions using the anisotropic analytic algorithm and pencil-beam convolution.

Authors:  Paola Francisca Caprile; Carlos Daniel Venencia; Pelayo Besa
Journal:  J Appl Clin Med Phys       Date:  2006-08-08       Impact factor: 2.102

  1 in total