Accuracy and clinical performance of a continuous intra-arterial blood-gas monitoring system during thoracoscopic surgery.

A Zollinger1, D R Spahn, T Singer, M P Zalunardo, S Stoehr, W Weder, T Pasch.   

Abstract

Accuracy and performance of the only currently available intra-arterial blood-gas monitoring system (Paratrend 7, PT7) were assessed in 23 patients during thoracoscopic surgery using one-lung ventilation. Over a wide range of values for arterial PO2 (6.1-61.1 kPa), PCO2 (4.1-9.5 kPa) and pH (7.19-7.50), 138 arterial blood-gas values obtained by PT7 were compared with corresponding in vitro laboratory blood-gas measurements. We found good clinical performance with the PT7 and good agreement between PT7 values and in vitro measurements for arterial PO2 (bias (1.96 SD) = 0.38 (9.52) kPa), PCO2 (0.31 (0.76) kPa) and pH (-0.017 (0.065)). Also, the bias for sequential changes between two, consecutive times was not significantly different from the ideal value of 0. We conclude that the PT7 is helpful in monitoring patients during thoracoscopy.

Entities:  

Mesh:

Substances:

Year:  1997        PMID: 9301388     DOI: 10.1093/bja/79.1.47

Source DB:  PubMed          Journal:  Br J Anaesth        ISSN: 0007-0912            Impact factor:   9.166


  3 in total

1.  Continuous neonatal blood gas monitoring using a multiparameter intra-arterial sensor.

Authors:  C Morgan; S J Newell; D A Ducker; J Hodgkinson; D K White; C J Morley; J M Church
Journal:  Arch Dis Child Fetal Neonatal Ed       Date:  1999-03       Impact factor: 5.747

2.  Comparison of local measurement of cerebral metabolism and to cerebral PvO2 during alterations in intracranial pressure, PaCO2 and arterial pressure--an experimental study in goat.

Authors:  Aram Ter Minassian; Jean Claude Desfontis; Freddy Gautier; Claire Douart; Mauro Ursino; Benoît Denizot; Marc Gogny; Laurent Beydon
Journal:  J Clin Monit Comput       Date:  2002-12       Impact factor: 2.502

3.  Assessment of a continuous blood gas monitoring system in animals during circulatory stress.

Authors:  Sandro Gelsomino; Roberto Lorusso; Ugolino Livi; Stefano Romagnoli; Salvatore Mario Romano; Rocco Carella; Fabiana Lucà; Giuseppe Billè; Francesco Matteucci; Attilio Renzulli; Gil Bolotin; Giuseppe De Cicco; Pierluigi Stefàno; Jos Maessen; Gian Franco Gensini
Journal:  BMC Anesthesiol       Date:  2011-01-11       Impact factor: 2.217