Comparative biomechanical analysis of supracondylar femur fracture fixation: locked intramedullary nail versus 95-degree angled plate.

S M David1, M E Harrow, R D Peindl, S L Frick, J F Kellam.   

Abstract

OBJECTIVES: To compare the initial stability of the genucephalic (GSH) intramedullary nail and the 95-degree condylar compression screw and side plate (DCS) for distal femur fractures.
DESIGN: Human cadaveric biomechanical study. PARTICIPANTS: Twelve matched pairs of fresh frozen human cadaveric femurs. INTERVENTION: Genucephalic intramedullary nail device (Smith and Nephew Richards, Memphis, TN, U.S.A.) and the 95-degree DCS device (Synthes USA, Paoli, PA, U.S.A.) were compared. Grouped or dispersed screw constructs were tested for each fracture fixation system with progressively more severe simulated fracture patterns. MAIN OUTCOME MEASUREMENT: Axial and torsional stiffness values.
RESULTS: The DCS plate with the dispersed screw configuration had the greatest torsional stiffness (p < 0.0011). The GSH nail with the grouped screw configuration absorbed more energy (work) during axial loading compared with the plate constructs (p < 0.0007). There were no significant differences in axial or torsional stiffness within treatment groups for fracture patterns of increasing severity.
CONCLUSIONS: Based on the authors' results, the selection of a GSH nail or a DCS plate should not be determined by the severity of the fracture. If a DCS plate construct is selected, the authors recommend a dispersed screw configuration, including the most proximal hole in the plate, to provide superior stiffness in torsional loading and equal stiffness in axial loading when compared with the GSH nail constructs. If a GSH nail is selected, the authors recommend a grouped screw configuration, which absorbed more energy during axial loading compared with the DCS plate constructs and the nail with the dispersed screw configuration.

Entities:  

Mesh:

Year:  1997        PMID: 9294798     DOI: 10.1097/00005131-199707000-00008

Source DB:  PubMed          Journal:  J Orthop Trauma        ISSN: 0890-5339            Impact factor:   2.512


  4 in total

1.  Nailing versus plating for comminuted fractures of the distal femur: a comparative biomechanical in vitro study of three implants.

Authors:  I Mehling; P Hoehle; W Sternstein; J Blum; P M Rommens
Journal:  Eur J Trauma Emerg Surg       Date:  2013-01-10       Impact factor: 3.693

2.  Biomechanical testing of distal femur osteotomy plate fixation techniques: the role of simulated physiological loading.

Authors:  Justus-Martijn Brinkman; Christof Hurschler; Jens Agneskirchner; Philip Lobenhoffer; René M Castelein; Ronald J van Heerwaarden
Journal:  J Exp Orthop       Date:  2014-06-26

3.  Biomechanical comparison of dynamic condylar screw and locking compression plate fixation in unstable distal femoral fractures: An in vitro study.

Authors:  Ashutosh Kumar Singh; Amit Rastogi; Vakil Singh
Journal:  Indian J Orthop       Date:  2013-11       Impact factor: 1.251

4.  FIXATION OF SUPRACONDYLAR FEMORAL FRACTURES: A BIOMECHANICAL ANALYSIS COMPARING 95° BLADE PLATES AND DYNAMIC CONDYLAR SCREWS (DCS).

Authors:  Marco Antônio Percope Andrade; André Soares Rodrigues; Celso Junio Mendonça; Luiz Gustavo Santos Portela
Journal:  Rev Bras Ortop       Date:  2015-11-16
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.