Literature DB >> 9293371

Intrinsic Dacron graft failure 19 years post-implantation.

D B Nunn1, B Pourdeyhimi.   

Abstract

Intrinsic failure of a knitted Dacron aortic bifurcation graft 19 years after implantation is reported. Failure was due to inability of the dilated and deteriorated left graft limb to hold anastomotic sutures. Numerous transverse cracks and breaks were detected in the Dacron filaments, in addition to small holes throughout the fabric structure, and a severe reduction in the yarn's tensile strength. Criteria are defined for establishing a correct diagnosis, and causative factors, incidence, and clinical management are discussed.

Entities:  

Mesh:

Substances:

Year:  1997        PMID: 9293371     DOI: 10.1016/s0967-2109(97)00028-8

Source DB:  PubMed          Journal:  Cardiovasc Surg        ISSN: 0967-2109


  4 in total

1.  Pseudoaneurysm in the Left Groin due to Ruptured Knitted Dacron Graft.

Authors:  Masahiro Ryugo; Takumi Yasugi; Mitsugi Nagashima; Hironori Izutani; Toru Okamura; Fumiaki Shikata; Masashi Kawamura; Kanji Kawachi
Journal:  Ann Vasc Dis       Date:  2011-06-02

2.  Clinical analysis of non-anastomotic aneurysms of implanted prosthetic grafts.

Authors:  Satoshi Yamamoto; Katsuyuki Hoshina; Hideo Kimura; Hiroyuki Okamoto; Kunihiro Shigematsu; Tetsuro Miyata; Toshiaki Watanabe
Journal:  Surg Today       Date:  2014-04-03       Impact factor: 2.549

3.  A Rare Case of Dacron Graft Rupture due to Friction against a Rib.

Authors:  Katsuhiko Matsuyama; Masahiko Kuinose; Nobusato Koizumi; Toru Iwahashi; Kayo Toguchi; Hitoshi Ogino
Journal:  Thorac Cardiovasc Surg Rep       Date:  2014-07-16

4.  A Unique Case of an Expanded Polytetrafluoroethylene Graft Rupture 14 Years after Abdominal Aortic Aneurysm Open Repair Managed with Placement of a Thoracic Endograft.

Authors:  Loukia Alexopoulou-Prounia; Stavros K Kakkos; Chrysanthi P Papageorgopoulou; Konstantinos Katsanos; Peter Zampakis; Polyzois Tsantrizos; Ioannis Ntouvas
Journal:  Aorta (Stamford)       Date:  2021-12-28
  4 in total