Miniature C-arm imaging: an in vitro study of detecting foreign bodies in the emergency department.

D M Cohen1, C T Garcia, A M Dietrich, R W Hickey.   

Abstract

STUDY
OBJECTIVE: Recently, we have had clinical success detecting foreign bodies (FBs) using a mobile C-arm fluoroscopic device. This study tests its utility to detect FBs of differing densities in soft tissue.
DESIGN: Blinded, randomized, controlled in vitro study.
METHODS: Two physicians used the Xi-scan mini C-arm to image FBs. Five FBs of differing densities were studied: metal, gravel, glass, wood, and plastic. The FBs were placed into the deep muscles of chicken legs. One hundred observations were made: 50 legs with FBs and 50 legs without FBs. The blinded investigators imaged the legs and determined the presence or absence of FBs.
RESULTS: Imaging located 100% of metal, gravel and glass FBs. Plastic and wood could not be consistently detected (sensitivity 0.4, specificity 0.6).
CONCLUSIONS: This device accurately detects metal, gravel and glass. Radiolucent (wood) and semiradiopaque (plastic) FBs could not be located reliably. Clinical trials would define utility of this device in saving time, money and radiation exposure.

Entities:  

Mesh:

Year:  1997        PMID: 9291509

Source DB:  PubMed          Journal:  Pediatr Emerg Care        ISSN: 0749-5161            Impact factor:   1.454


  5 in total

1.  Diagnostic and therapeutic aspects in the treatment of gunshot wounds of the viscerocranium.

Authors:  A Gröbe; J Klatt; M Heiland; R Schmelzle; P Pohlenz
Journal:  Eur J Trauma Emerg Surg       Date:  2010-05-11       Impact factor: 3.693

2.  A model for teaching bedside detection of glass in wounds.

Authors:  Matthew R Levine; Stephen M Gorman; Paul R Yarnold
Journal:  Emerg Med J       Date:  2007-06       Impact factor: 2.740

3.  Closed reduction internal fixation rates and procedure times for metacarpal fractures treated in a minor surgery area before and after the introduction of a mini C-arm unit.

Authors:  R C Mahabir; C M DeCroff; L Thurgood; A R Harrop
Journal:  Can J Plast Surg       Date:  2008

Review 4.  Needle breakage during local anesthesia in the oral cavity--a retrospective of the last 50 years with guidelines for treatment and prevention.

Authors:  Marcello Augello; Jeannette von Jackowski; Klaus Wilhelm Grätz; Christine Jacobsen
Journal:  Clin Oral Investig       Date:  2010-07-13       Impact factor: 3.573

5.  The use of navigation (BrainLAB Vector vision(2)) and intraoperative 3D imaging system (Siemens Arcadis Orbic 3D) in the treatment of gunshot wounds of the maxillofacial region.

Authors:  Alexander Gröbe; Christoph Weber; Rainer Schmelzle; Max Heiland; Jan Klatt; Philipp Pohlenz
Journal:  Oral Maxillofac Surg       Date:  2009-09
  5 in total