Literature DB >> 9284243

Monte Carlo calculation of dose rate distributions around 192Ir wires.

F Ballester1, C Hernández, J Pérez-Calatayud, F Lliso.   

Abstract

Monte Carlo calculations of absolute dose rate in liquid water are presented in the form of away-along tables for 1 and 5 cm 192Ir wires of 0.3 mm diameter. Simulated absolute dose rate values can be used as benchmark data to verify the calculation results of treatment planning systems or directly as input data for treatment planning. Best fit value of attenuation coefficient suitable for use in Sievert-integrals-type calculations has been derived based on Monte Carlo calculation results. For the treatment planning systems that are based on TG43 formalism we have also calculated the required dosimetry parameters.

Entities:  

Mesh:

Substances:

Year:  1997        PMID: 9284243     DOI: 10.1118/1.598142

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  6 in total

1.  A Monte Carlo study on dose distribution evaluation of Flexisource (192)Ir brachytherapy source.

Authors:  Majid Alizadeh; Mahdi Ghorbani; Abbas Haghparast; Naser Zare; Toktam Ahmadi Moghaddas
Journal:  Rep Pract Oncol Radiother       Date:  2015-02-23

2.  Dosimetric investigation of LDR brachytherapy ¹⁹²Ir wires by Monte Carlo and TPS calculations.

Authors:  Ahmet Bozkurt; Hediye Acun; Gonul Kemikler
Journal:  Jpn J Radiol       Date:  2012-10-10       Impact factor: 2.374

3.  Determination of exit skin dose for 192Ir intracavitary accelerated partial breast irradiation with thermoluminescent dosimeters.

Authors:  Julie A Raffi; Stephen D Davis; Cliff G Hammer; John A Micka; Keith A Kunugi; Jana E Musgrove; John W Winston; Terresa J Ricci-Ott; Larry A DeWerd
Journal:  Med Phys       Date:  2010-06       Impact factor: 4.071

4.  Computed organ doses to an Indian reference adult during brachytherapy treatment of esophagus, breast, and neck cancers.

Authors:  Biju Keshavkumar
Journal:  J Med Phys       Date:  2012-07

5.  HDR brachytherapy in vivo source position verification using a 2D diode array: A Monte Carlo study.

Authors:  Joel Poder; Dean Cutajar; Susanna Guatelli; Marco Petasecca; Andrew Howie; Joseph Bucci; Anatoly Rosenfeld
Journal:  J Appl Clin Med Phys       Date:  2018-06-01       Impact factor: 2.102

6.  Dosimetry of indigenously developed (192)Ir high-dose rate brachytherapy source: An EGSnrc Monte Carlo study.

Authors:  Sridhar Sahoo; T Palani Selvam; S D Sharma; Trupti Das; A C Dey; B N Patil; K V S Sastri
Journal:  J Med Phys       Date:  2016 Apr-Jun
  6 in total