Literature DB >> 9279669

Correlation between intraocular lens hydrophilicity and anterior capsule opacification and aqueous flare.

K Miyake1, I Ota, S Miyake, K Maekubo.   

Abstract

PURPOSE: To study the correlation between a basic parameter of intraocular lens biocompatibility, hydrophilicity, and two clinical parameters, postoperative inflammation and anterior capsule opacification.
SETTING: Miyake Eye Hospital, Nagoya, Japan.
METHODS: Three combinations of IOLs that were identical in shape but had distinct contact angles of water were used in this prospective double-masked study: (1) experimental comparison of collagen type IV and poly(methyl methacrylate) (PMMA) IOLs in rabbit eyes; (2) clinical comparison of heparin-surface-modified and PMMA IOLs; (3) clinical comparison of three foldable IOLs, silicone, acrylic, and memory. One of the two IOLs being compared in each situation was randomly assigned to both eyes of each animal or patient. At 1 and 3 months postoperatively, the degree of anterior capsule opacification and the amount of flare in the anterior chamber were determined.
RESULTS: In all three comparative situations, greater postoperative inflammation and more rapid anterior capsule opacification was seen in the eyes with hydrophobic IOLs.
CONCLUSION: There was a correlation between the hydrophilicity of an IOL and the severity of postoperative inflammation and the speed of anterior capsule opacification.

Entities:  

Mesh:

Substances:

Year:  1996        PMID: 9279669     DOI: 10.1016/s0886-3350(96)80159-3

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  11 in total

1.  Physical and cytological characters of carbon, titanium surface modified intraocular lens in rabbit eyes.

Authors:  Zhaoxu Yuan
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2003-08-23       Impact factor: 3.117

2.  Capsular contraction syndrome following insertion of hydrophilic acrylic lens.

Authors:  Archana Malik; Neeti Gupta; Sunandan Sood
Journal:  Int Ophthalmol       Date:  2011-01-29       Impact factor: 2.031

3.  A 1-year study on carbon, titanium surface-modified intraocular lens in rabbit eyes.

Authors:  Zhaoxu Yuan; Huimin Sun; Jiaqin Yuan
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2004-09-10       Impact factor: 3.117

4.  Influence of super-hydrophobic silicone rubber substrate on the growth and differentiation of human lens epithelial cells.

Authors:  Liuxueying Zhong; Jiayi Jin; Danying Zheng; Weisheng Guan; Yue Guo; Anfu Chen; Yujiang Peng; Qianying Gao; Yongxin Zheng; Hanxiong Huang
Journal:  J Mater Sci Mater Med       Date:  2018-11-13       Impact factor: 3.896

5.  Influence of intraocular lens material and design on postoperative intracapsular cellular reactivity.

Authors:  D J Apple
Journal:  Trans Am Ophthalmol Soc       Date:  2000

6.  Effect of the hydrophilicity of acrylic intraocular lens material and haptic angulation on anterior capsule opacification.

Authors:  Lorenz Vock; Michael Georgopoulos; Thomas Neumayer; Wolf Buehl; Oliver Findl
Journal:  Br J Ophthalmol       Date:  2006-11-15       Impact factor: 4.638

7.  Hydrophobic acrylic versus heparin surface-modified polymethylmethacrylate intraocular lens: a biocompatibility study.

Authors:  Daniele Tognetto; Lisa Toto; Daniela Minutola; Enzo Ballone; Marta Di Nicola; Rocco Di Mascio; Giuseppe Ravalico
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2003-07-17       Impact factor: 3.117

8.  Effect of haptic material and number of intraocular lens on anterior capsule contraction after cataract surgery.

Authors:  Sun Young Kim; Ji Wook Yang; Young Chun Lee; Su-Young Kim
Journal:  Korean J Ophthalmol       Date:  2013-01-15

9.  Uveal and capsular biocompatibility of two foldable acrylic intraocular lenses in patients with endogenous uveitis--a prospective randomized study.

Authors:  Martin Roesel; Carsten Heinz; Britta Heimes; Joerg Michael Koch; Arnd Heiligenhaus
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2008-07-19       Impact factor: 3.117

10.  Anterior capsule phimosis and capsular block syndrome in a patient with Steinert myotonic dystrophy: a case report.

Authors:  Nicola Rosa; Michele Lanza; Maddalena De Bernardo; Maria Borrelli; Luisa Politano
Journal:  Cases J       Date:  2009-12-10
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.