Literature DB >> 9276554

Risk of contamination from laparoscopic carbon dioxide insufflators.

G W Couper1, S W Ewen, Z H Krukowski.   

Abstract

Ten high flow laparoscopic carbon dioxides insufflators were examined to determine whether there was significant particulate or bacterial contamination of the gas delivered to patients. The gas delivery tubing and connections in the insufflators were also examined for the presence of bacterial colonization and for evidence of retrograde passage of body fluids to the insufflator. Metallic particulate debris was recovered from gas from all 10 insufflators. Staphylococci were recovered from the insufflated gas from one insufflator and from the internal tubing of three insufflators. Swabs from the internal tubing of two insufflators showed evidence of contamination by blood. The metallic particulate contamination of the insufflated gas may not be of clinical significance but the presence of bacteria and blood is a concern. Simple measures to minimize the risk of retrograde flow of gas and fluid to the insufflator should be followed and the provision of a filter between the patient and the insufflator is recommended.

Entities:  

Mesh:

Substances:

Year:  1997        PMID: 9276554

Source DB:  PubMed          Journal:  J R Coll Surg Edinb        ISSN: 0035-8835


  2 in total

1.  Current Evidence for Minimally Invasive Surgery During the COVID-19 Pandemic and Risk Mitigation Strategies: A Narrative Review.

Authors:  Sami A Chadi; Keegan Guidolin; Antonio Caycedo-Marulanda; Abdu Sharkawy; Antonino Spinelli; Fayez A Quereshy; Allan Okrainec
Journal:  Ann Surg       Date:  2020-05-20       Impact factor: 12.969

2.  Reduction of laparoscopic-induced hypothermia, postoperative pain and recovery room length of stay by pre-conditioning gas with the Insuflow device: a prospective randomized controlled multi-center study.

Authors:  D E Ott; H Reich; B Love; R McCorvey; A Toledo; C Y Liu; R Syed; K Kumar
Journal:  JSLS       Date:  1998 Oct-Dec       Impact factor: 2.172

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.