Literature DB >> 9270918

Automatic washer disinfector for flexible endoscopes: a new evaluation process.

L Pineau1, C Roques, J Luc, G Michel.   

Abstract

BACKGROUND AND STUDY AIMS: Many automatic washer disinfectors for flexible endoscopes have been marketed and offered as an alternative method of preventing infections, but they are frequently unsatisfactory. There is therefore clearly a need to test prototypes prior to marketing, following an evaluation process that is sufficiently reliable and rigorous to guarantee the efficacy of the decontamination processes.
MATERIALS AND METHODS: The present study describes an experimental method based on the follow-up of the decontamination of a Tygon tube, the internal surface of which was contaminated by a bacterial biofilm. This method is proposed as a preliminary test for evaluating washer disinfectors.
RESULTS: An analysis of the results obtained after technical modifications of the first prototype of the Fibro-Cleaner showed that complementary activities of each successive cycle phase allow a reduction in the number of adherent bacteria of more than 8 log per cm2 of support. With the three different biofilms tested (Escherichia coli, Pseudomonas aeruginosa mucoid and Staphylococcus aureus), no microorganisms were recovered from the support at the end of the decontamination process.
CONCLUSIONS: The experimental protocol suggested here seems to be well suited for assessing washer disinfectors during the development phase of the prototype, as well as for comparative studies.

Entities:  

Mesh:

Year:  1997        PMID: 9270918     DOI: 10.1055/s-2007-1004218

Source DB:  PubMed          Journal:  Endoscopy        ISSN: 0013-726X            Impact factor:   10.093


  4 in total

1.  Interactions between biocide cationic agents and bacterial biofilms.

Authors:  C Campanac; L Pineau; A Payard; G Baziard-Mouysset; C Roques
Journal:  Antimicrob Agents Chemother       Date:  2002-05       Impact factor: 5.191

2.  Assessment on experimental bacterial biofilms and in clinical practice of the efficacy of sampling solutions for microbiological testing of endoscopes.

Authors:  C Aumeran; E Thibert; F A Chapelle; C Hennequin; O Lesens; O Traoré
Journal:  J Clin Microbiol       Date:  2011-12-14       Impact factor: 5.948

Review 3.  Is peracetic acid suitable for the cleaning step of reprocessing flexible endoscopes?

Authors:  Günter Kampf; Patricia M Fliss; Heike Martiny
Journal:  World J Gastrointest Endosc       Date:  2014-09-16

4.  Activity of ethanol and daptomycin lock on biofilm generated by an in vitro dynamic model using real subcutaneous injection ports.

Authors:  C Aumeran; P Guyot; M Boisnoir; C Robin-Hennequin; M Vidal; C Forestier; O Traore; O Lesens
Journal:  Eur J Clin Microbiol Infect Dis       Date:  2012-10-19       Impact factor: 3.267

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.