Literature DB >> 9258493

Successful management of Palmaz-Schatz stents deformed by inadvertent angioplasty outside the stents.

T N Chu1, F S Ling.   

Abstract

We describe a case in which balloon angioplasty of thrombosed saphenous vein graft stents resulted in compression and deformation of the stents because of inadvertent passage of the guidewire outside the stents. This condition was diagnosed by intravascular ultrasonography, which was then ultimately utilized to direct and optimize placement of additional stents to exclude the deformed stents from the vessel lumen.

Mesh:

Substances:

Year:  1997        PMID: 9258493     DOI: 10.1002/(sici)1097-0304(199708)41:4<435::aid-ccd22>3.0.co;2-j

Source DB:  PubMed          Journal:  Cathet Cardiovasc Diagn        ISSN: 0098-6569