Effect of cataract and pupil size on image quality with confocal scanning laser ophthalmoscopy.

L Zangwill1, I Irak, C C Berry, V Garden, M de Souza Lima, R N Weinreb.   

Abstract

OBJECTIVE: To determine the effect of pupil size and cataract on the reproducibility and image quality obtained with confocal scanning laser ophthalmoscopy.
METHODS: Three image series were obtained with a confocal scanning laser ophthalmoscope (Heidelberg Retina Tomograph, Heidelberg, Germany) before and after pupillary dilation in each of 39 subjects (8 normal subjects, 5 glaucoma suspects, and 26 patients with glaucoma). The cataract density was measured with both a lens opacity meter and the Lens Opacities Classification System III system. The image quality of each image series was subjectively scored on a scale of 0 (poor) to 9 (high) by 2 independent observers who were unaware of the cataract density and pupil size during image acquisition. The image quality and reproducibility were objectively evaluated using the SD of the mean topography image of each subject.
RESULTS: The mean pupil diameter in all subjects before and after dilation was 2.5 +/- 0.8 mm and 5.8 +/- 1.4 mm, respectively. After pupillary dilation, both the mean image quality score and mean SD of the mean topography image improved (from 4.5 +/- 3.5 to 7.4 +/- 2.3 and from 48.6 +/- 18.8 microns to 35.6 +/- 15.5 microns, respectively). The quality score of the images that were obtained before pupillary dilation was associated with the pupil size and density of nuclear and posterior subcapsular cataracts. The SD of the mean topography images that were obtained before pupillary dilation increased with a decreasing pupil size (P = .003) and an increasing density of the nuclear (P < .03), cortical (P = .02), and posterior subcapsular (P = .002) opacity.
CONCLUSIONS: Although pupillary dilation improved the image quality in most subjects, the improvement was sometimes small. Those subjects with small undilated pupils and/or cataracts may benefit most from pupillary dilation.

Entities:  

Mesh:

Year:  1997        PMID: 9258219     DOI: 10.1001/archopht.1997.01100160153003

Source DB:  PubMed          Journal:  Arch Ophthalmol        ISSN: 0003-9950


  10 in total

1.  Optic nerve head morphometry in healthy adults using confocal laser scanning tomography.

Authors:  M M Hermann; I Theofylaktopoulos; N Bangard; C Jonescu-Cuypers; S Coburger; M Diestelhorst
Journal:  Br J Ophthalmol       Date:  2004-06       Impact factor: 4.638

2.  No changes in anatomical and functional glaucoma evaluation after trabeculectomy.

Authors:  Ivan Maynart Tavares; Luiz Alberto S Melo; João A Prata; Roberta Galhardo; Augusto Paranhos; Paulo Augusto A Mello
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2005-09-01       Impact factor: 3.117

3.  Detecting glaucoma with RADAAR: the Bridlington Eye Assessment Project.

Authors:  M J Hawker; S A Vernon; C L Tattersall; H S Dua
Journal:  Br J Ophthalmol       Date:  2006-06       Impact factor: 4.638

4.  Effect of pupillary dilatation on glaucoma assessments using optical coherence tomography.

Authors:  M Smith; A Frost; C M Graham; S Shaw
Journal:  Br J Ophthalmol       Date:  2007-06-07       Impact factor: 4.638

5.  Some dissociating factors in the analysis of structural and functional progressive damage in open-angle glaucoma.

Authors:  C J W Hudson; L S Kim; S A Hancock; I A Cunliffe; J M Wild
Journal:  Br J Ophthalmol       Date:  2006-10-18       Impact factor: 4.638

6.  Influence of cataract on optical coherence tomography image quality and retinal thickness.

Authors:  M E J van Velthoven; M H van der Linden; M D de Smet; D J Faber; F D Verbraak
Journal:  Br J Ophthalmol       Date:  2006-10       Impact factor: 4.638

7.  Factors affecting the test-retest variability of Heidelberg retina tomograph and Heidelberg retina tomograph II measurements.

Authors:  N G Strouthidis; E T White; V M F Owen; T A Ho; C J Hammond; D F Garway-Heath
Journal:  Br J Ophthalmol       Date:  2005-11       Impact factor: 4.638

8.  Scan quality effect on glaucoma discrimination by glaucoma imaging devices.

Authors:  K R Sung; G Wollstein; J S Schuman; R A Bilonick; H Ishikawa; K A Townsend; L Kagemann; M L Gabriele
Journal:  Br J Ophthalmol       Date:  2009-08-18       Impact factor: 4.638

Review 9.  Imaging systems of human eye: a review.

Authors:  Rajendra U Acharya; Wong Li Yun; E Y K Ng; Wenwei Yu; Jasjit S Suri
Journal:  J Med Syst       Date:  2008-08       Impact factor: 4.460

10.  Correlation between morphology of optic disc determined by Heidelberg Retina Tomograph II and visual function in eyes with open-angle glaucoma.

Authors:  Kazuko Omodaka; Toru Nakazawa; Takaaki Otomo; Masahiko Nakamura; Nobuo Fuse; Kohji Nishida
Journal:  Clin Ophthalmol       Date:  2010-07-30
  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.