A pressure-balanced radio-telemetry system for the measurement of intracranial pressure. A preliminary design report.

N T Zervas, E R Cosman, B J Cosman.   

Abstract

A fully implantable radio-telemetric differential intracranial pressure sensor is described with a zero-point calibration that can be confirmed repeatedly. Intracranial pressure (ICP) is measured with the device by the principle of applying a known external pressure to the scalp above the sensor and simultaneously detecting by radio-telemetry the zero-point of the sensor corresponding to a balance of pressures across it. The radio-telemetry is implemented by a resonant circuit in the sensor of which the resonant radiofrequency is detected outside the body. The sensor is passive, has built-in barametric compensation, negligible permeability of temperature drift, no calibration ambiguities, and fast dynamic response. The implanted sensor has been used successfully for short-term as well as long-term ICP monitoring. It has been implemented primarily for intermittent ICP measurements, but also adapted to continuous recording. Preliminary clinical experience with the system indicates that it is effective, safe, and simple to operate.

Mesh:

Year:  1977        PMID: 925744     DOI: 10.3171/jns.1977.47.6.0899

Source DB:  PubMed          Journal:  J Neurosurg        ISSN: 0022-3085            Impact factor:   5.115


  6 in total

1.  Detection of hidden pseudotumour cerebri behind Chiari 1 malformation: value of telemetric ICP monitoring.

Authors:  Melanie Schmitt; Michael Kiefer; Sebastian Antes; Regina Eymann
Journal:  Childs Nerv Syst       Date:  2012-04-20       Impact factor: 1.475

2.  Otogenic intracranial abscesses.

Authors:  A Kulai; N Ozatik; I Topçu
Journal:  Acta Neurochir (Wien)       Date:  1990       Impact factor: 2.216

3.  Technique for very long-term monitoring of intracranial pressure.

Authors:  G S Brindley; C E Polkey; J D Cooper
Journal:  Med Biol Eng Comput       Date:  1983-07       Impact factor: 2.602

4.  Batteryless on-demand-sampling active radiosonde for intracranial pressure measurement.

Authors:  V Macellari
Journal:  Med Biol Eng Comput       Date:  1981-11       Impact factor: 2.602

5.  The Rotterdam Teletransducer: state of the device.

Authors:  A I Maas; D A de Jong
Journal:  Acta Neurochir (Wien)       Date:  1986       Impact factor: 2.216

Review 6.  Advances in Intracranial Pressure Monitoring and Its Significance in Managing Traumatic Brain Injury.

Authors:  Usmah Kawoos; Richard M McCarron; Charles R Auker; Mikulas Chavko
Journal:  Int J Mol Sci       Date:  2015-12-04       Impact factor: 5.923

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.