| Literature DB >> 9253814 |
S D Pathak1, V Chalana, Y Kim.
Abstract
We have developed a tool to automatically detect inner and outer skull boundaries of a fetal head in ultrasound images. These boundaries are used to measure biparietal diameter (BPD) and head circumference (HC). The algorithm is based on active contour models and takes 32 s on a Sun SparcStation 20/71. A high-performance desktop multimedia system called MediaStation 5000 (MS5000) is used as a model for our future ultrasound subsystem. On the MS5000, the optimized implementation of this algorithm takes 248 ms. The difference (between the computer-measured values on MS5000 and the gold standard) for BPD and HC was 1.43% (sigma = 1.00%) and 1.96% (sigma = 1.96%), respectively. According to our data analysis, no significant differences exist in the BPD and HC measurements made on the MS5000 and those measurements made on the Sun SparcStation 20/71. Reduction in the overall execution time from 32 s to 248 ms will help making this algorithm a practical ultrasound tool for sonographers.Mesh:
Year: 1997 PMID: 9253814 DOI: 10.1016/s0301-5629(97)00009-4