The optimum intensities for multiple static multileaf collimator field compensation.

P M Evans1, V N Hansen, W Swindell.   

Abstract

A method of determining the optimum beam intensities for compensation using multiple static multileaf collimator fields is presented. In this method a histogram of the number of beam pixels against beam intensity is generated for the intensity-modulated beam (IMB). The intensity of each beam to be used is chosen to minimize the mean square deviation between each bin in the histogram and the closest beam intensity. This method has been applied to sample IMBs possessing one maximum and two maxima. For both cases, the use of uniform beam intensity increments is shown to be close to optimal. In the case with two maxima, the efficacy of irradiating both peaks simultaneously, rather than separately, has been studied and shown to be of potential benefit. The optimum intensities for an IMB for breast radiotherapy are also presented.