Quality assurance system to correct for errors arising from couch rotation in linac-based stereotactic radiosurgery.

I A Brezovich1, P N Pareek, W E Plott, R L Jennelle.   

Abstract

PURPOSE: The purpose of this project was the development of a quality assurance (QA) system that would provide geographically accurate targeting for linac-based stereotactic radiosurgery (LBSR). METHODS AND MATERIALS: The key component of our QA system is a novel device (Alignment Tool) for expedient measurement of gantry and treatment table excursions (wobble) during rotation. The Alignment Tool replaces the familiar pencil-shaped pointers with a ball pointer that is used with the field light of the accelerator to indicate alignment of beam and target. Wobble is measured prior to each patient treatment and analyzed together with the BRW coordinates of the target by a spreadsheet. The corrections required to compensate for any imprecisions are identified, and a printout generated indicating the floor stand coordinates for each couch angle used to place the target at isocenter.
RESULTS: The Alignment Tool has an inherent accuracy of measurement better than 0.1 mm. The overall targeting error of our QA method, found by evaluating 177 target simulator films of 55 foci in 40 randomly selected patients, was 0.47 +/- 0.23 mm. The Alignment Tool was also valuable during installation of the floor stand and a supplemental collimator for the accelerator.
CONCLUSIONS: The QA procedure described allows accurate targeting in LBSR, even when couch rotation is imprecise. The Alignment Tool can facilitate the installation of any stereotactic irradiation system, and can be useful for annual QA checks as well as in the installation and commissioning of new accelerators.

Entities:  

Mesh:

Year:  1997        PMID: 9240658     DOI: 10.1016/s0360-3016(97)00057-6

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  3 in total

1.  A 3D star shot to determine the gantry, collimator, and couch axes positions.

Authors:  Robert Corns; Kaida Yang; Mason Ross; Shiva Bhandari; Makunda Aryal; Peter Ciaccio
Journal:  J Appl Clin Med Phys       Date:  2022-04-29       Impact factor: 2.243

Review 2.  Isocenter verification for linac-based stereotactic radiation therapy: review of principles and techniques.

Authors:  Pejman Rowshanfarzad; Mahsheed Sabet; Daryl J O'Connor; Peter B Greer
Journal:  J Appl Clin Med Phys       Date:  2011-11-15       Impact factor: 2.102

Review 3.  Does Stereotactic Radiosurgery Have a Role in the Management of Patients Presenting With 4 or More Brain Metastases?

Authors:  Michael H Soike; Ryan T Hughes; Michael Farris; Emory R McTyre; Christina K Cramer; J D Bourland; Michael D Chan
Journal:  Neurosurgery       Date:  2019-03-01       Impact factor: 5.315

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.