Literature DB >> 9236706

Long-term in vivo alterations of polyester vascular grafts in humans.

G Riepe1, J Loos, H Imig, A Schröder, E Schneider, J Petermann, A Rogge, M Ludwig, A Schenke, R Nassutt, N Chakfe, M Morlock.   

Abstract

OBJECTIVES: To examine the influence of in vivo hydrolysis on the physical properties of polyester grafts and their correlation to the period of implantation in the human body.
MATERIALS AND METHODS: Sixty-five explanted vascular grafts were obtained after 0-23 years of implantation due to suture aneurysms (18), occlusion (12), graft infection (12), failure of graft material (7) and post-mortem (16). The surface was examined by scanning electron microscopy, the molecular integrity by infra-red spectroscopy and physical strength by probe puncture.
RESULTS: Scission of macromolecular chains and loss of strength were shown. It was demonstrated that hydrolytic degradation of polyester takes place with increasing time of implantation in humans. Analysis by linear regression showed that polyester grafts lose 31.4% of their bursting strength in 10 years and 100% in 25-39 years after implantation.
CONCLUSIONS: Regular follow-ups of patients with aged vascular grafts and the precise documentation of implanted materials are necessary to estimate graft degradation.

Entities:  

Mesh:

Substances:

Year:  1997        PMID: 9236706     DOI: 10.1016/s1078-5884(97)80062-7

Source DB:  PubMed          Journal:  Eur J Vasc Endovasc Surg        ISSN: 1078-5884            Impact factor:   7.069


  19 in total

Review 1.  Design and features of the Acorn CorCap Cardiac Support Device: the concept of passive mechanical diastolic support.

Authors:  Robert G Walsh
Journal:  Heart Fail Rev       Date:  2005-06       Impact factor: 4.214

2.  [Treatment of abdominal wall defects, including abdominal relaxation].

Authors:  C Langer; H Becker
Journal:  Chirurg       Date:  2006-05       Impact factor: 0.955

Review 3.  Cardiovascular Bio-Engineering: Current State of the Art.

Authors:  Teresa Simon-Yarza; Isabelle Bataille; Didier Letourneur
Journal:  J Cardiovasc Transl Res       Date:  2017-03-06       Impact factor: 4.132

Review 4.  Mesh biocompatibility: effects of cellular inflammation and tissue remodelling.

Authors:  Karsten Junge; Marcel Binnebösel; Klaus T von Trotha; Raphael Rosch; Uwe Klinge; Ulf P Neumann; Petra Lynen Jansen
Journal:  Langenbecks Arch Surg       Date:  2011-04-01       Impact factor: 3.445

Review 5.  Biocompatibility of prosthetic meshes in abdominal surgery.

Authors:  Marcel Binnebösel; Klaus T von Trotha; Petra Lynen Jansen; Joachim Conze; Ulf P Neumann; Karsten Junge
Journal:  Semin Immunopathol       Date:  2011-01-12       Impact factor: 9.623

6.  Re: Totally extraperitoneal repair of inguinal hernias: some remarks on technical details.

Authors:  Christophe Roger Berney
Journal:  Langenbecks Arch Surg       Date:  2013-02-03       Impact factor: 3.445

Review 7.  Surgical perspectives regarding application of biomaterials for the management of large congenital diaphragmatic hernia defects.

Authors:  Amulya K Saxena
Journal:  Pediatr Surg Int       Date:  2018-04-02       Impact factor: 1.827

8.  Major mesh-related complications following hernia repair: events reported to the Food and Drug Administration.

Authors:  T N Robinson; J H Clarke; J Schoen; M D Walsh
Journal:  Surg Endosc       Date:  2005-10-05       Impact factor: 4.584

9.  In vivo biostability of polymeric spine implants: retrieval analyses from a United States investigational device exemption study.

Authors:  Ming Shen; Kai Zhang; Petra Koettig; William C Welch; John M Dawson
Journal:  Eur Spine J       Date:  2011-05-03       Impact factor: 3.134

10.  Materials characterization and histological analysis of explanted polypropylene, PTFE, and PET hernia meshes from an individual patient.

Authors:  A J Wood; M J Cozad; D A Grant; A M Ostdiek; S L Bachman; S A Grant
Journal:  J Mater Sci Mater Med       Date:  2013-01-31       Impact factor: 3.896