Literature DB >> 9214782

Magnetoencephalography with diversely oriented and multicomponent sensors.

B Hochwald1, A Nehorai.   

Abstract

To locate endocranial current sources, a magnetoencephalography (MEG) system usually measures the magnetic field at many points around the skull with an array of radial sensors. Despite the success of using radial components of the field, we show that using nonradial components may potentially also be beneficial. We demonstrate some benefits of using diversely oriented and multicomponent sensors to measure the nonradial components. A framework is provided for analyzing the accuracy of a system that estimates the location and direction of a current dipole inside a spherical skull. The framework is then used to determine the effect on accuracy of varying the orientations of sensors in an array and, as a consequence, it is found that the radial orientations commonly used in practice are suboptimal for locating dipoles near the array's center. A diversely oriented array that improves performance is presented. We show how a single multicomponent sensor can locate a dipole, and derive a simple algorithm for locating a dipole near the sensor.

Mesh:

Year:  1997        PMID: 9214782     DOI: 10.1109/10.553711

Source DB:  PubMed          Journal:  IEEE Trans Biomed Eng        ISSN: 0018-9294            Impact factor:   4.538


  4 in total

1.  Performance Analysis of Optically Pumped 4He Magnetometers vs. Conventional SQUIDs: From Adult to Infant Head Models.

Authors:  Saeed Zahran; Mahdi Mahmoudzadeh; Fabrice Wallois; Nacim Betrouni; Philippe Derambure; Matthieu Le Prado; Agustin Palacios-Laloy; Etienne Labyt
Journal:  Sensors (Basel)       Date:  2022-04-18       Impact factor: 3.847

2.  Optimal Magnetic Sensor Vests for Cardiac Source Imaging.

Authors:  Stephan Lau; Bojana Petković; Jens Haueisen
Journal:  Sensors (Basel)       Date:  2016-05-24       Impact factor: 3.576

3.  Improved spatio-temporal measurements of visually evoked fields using optically-pumped magnetometers.

Authors:  Aikaterini Gialopsou; Christopher Abel; T M James; Thomas Coussens; Mark G Bason; Reuben Puddy; Francesco Di Lorenzo; Katharina Rolfs; Jens Voigt; Tilmann Sander; Mara Cercignani; Peter Krüger
Journal:  Sci Rep       Date:  2021-11-17       Impact factor: 4.379

4.  Optimal design of on-scalp electromagnetic sensor arrays for brain source localisation.

Authors:  Leandro Beltrachini; Nicolas von Ellenrieder; Roland Eichardt; Jens Haueisen
Journal:  Hum Brain Mapp       Date:  2021-07-10       Impact factor: 5.038

  4 in total