| Literature DB >> 9212935 |
T Mussivand1, A Hum, K S Holmes, W J Keon.
Abstract
A wireless biotelemetry system for the transfer of digital data through intact skin and tissue has been developed to provide a safe and noninvasive means of communication between implanted medical devices and the outside of the body. The system utilizes 2 miniature infrared transmitter/receiver modules. Data are transmitted through intact skin and subcutaneous tissue on an 890 nm infrared carrier signal. The system has been evaluated in human cadavers and during in vivo implantation of artificial hearts and ventricular assist devices for durations of up to 96 h. Acceptable data transfer (error rate < 10(-5)) through a typical tissue thickness of 5-25 mm has been demonstrated. The ability to monitor and control a device from a remote site using public communication systems such as telephone lines and asynchronous transfer mode (ATM) systems has also been demonstrated. Design optimization is currently ongoing in preparation for clinical utilization with artificial heart systems and other implantable devices (such as rotary blood pumps).Entities:
Mesh:
Year: 1997 PMID: 9212935 DOI: 10.1111/j.1525-1594.1997.tb03718.x
Source DB: PubMed Journal: Artif Organs ISSN: 0160-564X Impact factor: 3.094