Silicone-coated polypropylene hollow-fiber oxygenator: experimental evaluation and preliminary clinical use.

T Shimono1, Y Shomura, I Hioki, A Shimamoto, H Tenpaku, Y Maze, K Onoda, M Takao, H Shimpo, I Yada.   

Abstract

BACKGROUND: A membrane oxygenator consisting of a microporous polypropylene hollow fiber with a 0.2-microm ultrathin silicone layer (cyclosiloxane) was developed. Animal experimental and preliminary clinical studies evaluated its reliability in bypass procedures.
METHODS: Five 24-hour venoarterial bypass periods were conducted on dogs using the oxygenator (group A). In 5 controls, bypass periods were conducted using the same oxygenator without silicone coating (group B). As a preliminary clinical study, 14 patients underwent cardiopulmonary bypass with the silicone-coated oxygenator.
RESULTS: Eight to 16 hours (mean, 12.2 hours) after initiation of bypass, plasma leakage occurred in all group B animals, but none in group A. The O2 and CO2 transfer rates after 24 hours in group A were significantly higher than at termination of bypass in group B (p < 0.005 and p < 0.03, respectively). Scanning electron microscopy of silicone-coated fibers after 24 hours of bypass revealed no damage to the silicone coating of the polypropylene hollow fibers. In the clinical study, the oxygenator showed good gas transfer, acceptable pressure loss, low hemolysis, and good durability.
CONCLUSIONS: This oxygenator is more durable and offers greater gas transfer capabilities than the previous generation of oxygenators.

Entities:  

Mesh:

Substances:

Year:  1997        PMID: 9205175     DOI: 10.1016/s0003-4975(97)00119-7

Source DB:  PubMed          Journal:  Ann Thorac Surg        ISSN: 0003-4975            Impact factor:   4.330


  3 in total

1.  Hemolysis-associated endothelial dysfunction mediated by accelerated NO inactivation by decompartmentalized oxyhemoglobin.

Authors:  Peter C Minneci; Katherine J Deans; Huang Zhi; Peter S T Yuen; Robert A Star; Steven M Banks; Alan N Schechter; Charles Natanson; Mark T Gladwin; Steven B Solomon
Journal:  J Clin Invest       Date:  2005-11-17       Impact factor: 14.808

2.  Gas Transfer in Cellularized Collagen-Membrane Gas Exchange Devices.

Authors:  Justin H Lo; Erik K Bassett; Elliot J N Penson; David M Hoganson; Joseph P Vacanti
Journal:  Tissue Eng Part A       Date:  2015-07-16       Impact factor: 3.845

3.  Hollow Fiber Oxygenator Composition Has a Significant Impact on Failure Rates in Neonates on Extracorporeal Membrane Oxygenation: A Retrospective Analysis.

Authors:  John M Daniel; Philip A Bernard; Sean C Skinner; Prasad Bhandary; Ana Ruzic; Matthew K Bacon; Hubert O Ballard
Journal:  J Pediatr Intensive Care       Date:  2017-03-07
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.