Abstract
Fixed-separation plane-parallel ionization chambers have been shown to overestimate the dose in the buildup region of normally incident high-energy photon beams. This work shows that these ionization chambers exhibit an even greater over-response in the buildup region of obliquely incident photon beams. This over-response at oblique incidence is greatest at the surface of the phantom and increases with increasing angle of beam incidence. In addition, the magnitude of the over-response depends on field size, beam energy, and chamber construction. This study shows that plane-parallel ionization chambers can over-respond by more than a factor of 2.3 at the phantom surface for obliquely incident high-energy photon fields.Entities:
Mesh:
Year: 1997 PMID: 9198021 DOI: 10.1118/1.598000
Source DB: PubMed Journal: Med Phys ISSN: 0094-2405 Impact factor: 4.071