Plane-parallel ionization chamber response in the buildup region of obliquely incident photon beams.

B J Gerbi1, F M Khan.   

Abstract

Fixed-separation plane-parallel ionization chambers have been shown to overestimate the dose in the buildup region of normally incident high-energy photon beams. This work shows that these ionization chambers exhibit an even greater over-response in the buildup region of obliquely incident photon beams. This over-response at oblique incidence is greatest at the surface of the phantom and increases with increasing angle of beam incidence. In addition, the magnitude of the over-response depends on field size, beam energy, and chamber construction. This study shows that plane-parallel ionization chambers can over-respond by more than a factor of 2.3 at the phantom surface for obliquely incident high-energy photon fields.

Entities:  

Mesh:

Year:  1997        PMID: 9198021     DOI: 10.1118/1.598000

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  3 in total

1.  Dose discrepancies in the buildup region and their impact on dose calculations for IMRT fields.

Authors:  Shu-Hui Hsu; Jean M Moran; Yu Chen; Ravi Kulasekere; Peter L Roberson
Journal:  Med Phys       Date:  2010-05       Impact factor: 4.071

2.  Surface dose measurements with commonly used detectors: a consistent thickness correction method.

Authors:  Tatsiana A Reynolds; Patrick Higgins
Journal:  J Appl Clin Med Phys       Date:  2015-09-08       Impact factor: 2.102

3.  A study of surface dosimetry for breast cancer radiotherapy treatments using Gafchromic EBT2 film.

Authors:  Masahiro Nakano; Robin F Hill; May Whitaker; Jung-Ha Kim; Zdenka Kuncic
Journal:  J Appl Clin Med Phys       Date:  2012-05-10       Impact factor: 2.102

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.