P Bean1, K Liegmann, T Løvli, C Westby, E Sundrehagen.
Abstract
Carbohydrate-deficient transferrin (CDT) may now be the most valuable biological marker for diagnosis of alcohol abuse. We compared the diagnostic performance of two new CDT tests, Axis %CDT turbidimetric immunoassay (TIA) and Axis %CDT HPLC, against Specialty Laboratories' isoelectric focusing/immunoblotting/laser densitometry (IEF/IB/LD). Both Axis tests include one-half the concentration of trisialotransferrin isoforms in their CDT quantitation schemes. Considering an alcohol abuse prevalence of 7%, Axis %CDT TIA shows a sensitivity of 87% at 98% specificity and a positive predictive value (PPV) of 0.75; %CDT HPLC shows a sensitivity of 87% at 100% specificity for a PPV of 1, and the IEF/IB/LD shows 81% sensitivity at 94% specificity for a PPV of 0.5. All three CDT tests show the same negative predictive value (0.98). Both Axis procedures perform better than IEF/IB/LD in the diagnosis of alcohol abuse; %CDT TIA is available in several semiautomated, cost-effective formats.Entities:
Mesh:
Substances:
Year: 1997 PMID: 9191550
Source DB: PubMed Journal: Clin Chem ISSN: 0009-9147 Impact factor: 8.327