Sound spectroanalytic diagnosis of malfunctioning prosthetic heart valve.

Y Kagawa, S Nitta, N Satoh, K Saji, Y Shibota.   

Abstract

Sound spectroanalysis was carried out for diagnosis of malfunctioning prosthetic heart valves in 50 cases, which consisted of 31 mitral, 14 aortic, 4 mitral+aortic and 1 tricuspid valve replacements with the age ranging from 17 to 57 years. Types of prosthetic valves used in this series were Starr-Edwards silicon ball valves, metal ball valves, disc valves, Kay-Shiley valves, Smeloff-Cutter valves and a Björk-Shiley valve. The system used in this study consisted of a moving coil microphone, a preamplifier, a filter (160/24 db octave) and a sound spectrograph (Rion Co. S.G. (7). Contour sound spectrograms (SSG) were drawn and peak frequency, wave pattern and frequency-intensity curves were investigated as characteristics of implanted valvular function. Peak frequencies of valvular opening and closing sounds were recorded between 1800 to 5000 Hz and differed according to the types of prosthesis. In mitral+aortic valve replacements, valvular sounds of two prostheses were clearly separated and this finding was very useful for determining the prosthesis which fell into malfunction. Decrease in peak frequency of valvular sounds or abnormal wave patterns in contour SSG were observed in 9 cases (18%). Of 9 cases, 4 cases underwent re-replacement and 2 cases died of congestive heart failure. Operative or autopsy findings gave proof to the findings in sound spectroanalysis.

Mesh:

Year:  1977        PMID: 918976     DOI: 10.1620/tjem.123.77

Source DB:  PubMed          Journal:  Tohoku J Exp Med        ISSN: 0040-8727            Impact factor:   1.848


  4 in total

1.  New automated wavelet analytical system with a cellular phone for recording intercellular phone remote transmitted bileaflet valve sound.

Authors:  Takashi Sugiki; Hiroshi Sugiki; Norihiko Shiiya; Yoshiro Matsui; Kenji Sugiki
Journal:  J Artif Organs       Date:  2009-12-25       Impact factor: 1.731

2.  Wavelet analysis of valve closing sound detects malfunction of bileaflet mechanical valve.

Authors:  Hiroshi Sugiki; Toshifumi Murashita; Norihiko Shiiya; Yoshiro Matsui; Kenji Sugiki
Journal:  J Artif Organs       Date:  2008-04-15       Impact factor: 1.731

3.  Frequency analysis of Ionescu-Shiley prosthetic closing sounds in patients with normally functioning prostheses.

Authors:  M Brais; L G Durand; M Blanchard; J de Guise; R Guardo; W J Keon
Journal:  Med Biol Eng Comput       Date:  1986-11       Impact factor: 2.602

4.  Time-frequency analysis of the first heart sound: Part 3: Application to dogs with varying cardiac contractility and to patients with mitral mechanical prosthetic heart valves.

Authors:  D Chen; L G Durand; H C Lee; D W Wieting
Journal:  Med Biol Eng Comput       Date:  1997-09       Impact factor: 2.602

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.