Development of a recording system for saliva pH with and complete denture by telemetry.

T Watanabe1, K Kobayashi, T Suzuki, H Kumagai, K Ichikawa, K Kobayashi, T Kyu, I Sato, M Nagao.   

Abstract

A new system for measuring saliva pH by radio telemetry was developed. The system was built into a lower complete denture and could operate for about 19 h. The maximum error in the range pH 5-8 was 0.15 pH, and the correlation when values were compared with those from a glass-electrode pH meter was 0.99. That this system operates successfully was confirmed in a clinical experiment.

Mesh:

Year:  1997        PMID: 9188996     DOI: 10.1016/s0003-9969(96)00119-7

Source DB:  PubMed          Journal:  Arch Oral Biol        ISSN: 0003-9969            Impact factor:   2.633