Literature DB >> 9186107

The perils of permanent punctal plugs.

C N Soparkar1, J R Patrinely, J Hunts, J V Linberg, R C Kersten, R Anderson.   

Abstract

PURPOSE: To describe previously unreported complications associated with permanent lacrimal punctal plugs.
METHOD: Five oculoplastic practices reviewed patients referred to them over the preceding 2 years for permanent lacrimal punctal plug complications.
RESULTS: In 12 patients, 14 lacrimal punctal plugs migrated distally within the lacrimal drainage system, causing symptoms and necessitating surgical removal.
CONCLUSION: Luxation of permanent punctal plugs into the distal lacrimal drainage system can occur, sometimes requiring complex surgical intervention.