Mechanical testing of the tension band wire fixation in the proximal femur.

E E Engel1, J B Volpon, A C Shimano.   

Abstract

The mechanical stability of proximal femoral osteotomies fixed by the tension band wire technique was studied in flexion-compression and torsion tests. The fixation consisted in crossing the section with two Kirschner wires and with a wire cerclage applied to the tension surface. The study was conducted in three steps. First, cyclinders of wood were cut either transversely or at 30 degrees of inclination in relation to the long axis of the specimen, and fixed with two Kirschner wires and a wire cerclage. We concluded that the inclination of the plane of section significantly increased the stability of fixation. No significant difference was observed when oblique sections were made in the reverse orientation. Second, 30 degrees subtrochanteric varus osteotomies were performed in dog femurs, so that the section plane was transverse in one group and oblique in another, after closing the osteotomy. In both groups the fixation was achieved by two Kirschner wires that crossed the osteotomy and a wire cerclage placed on the lateral cortex (tension surface). We concluded that inclination of the osteotomy plane increased the stability of osteosynthesis in bone specimens, as already seen with the wood pieces. Third, the stability of tension band wire fixation was compared with that provided by the AO/ASIF paediatric angled plate. Varus osteotomies (30 degrees) were created at the subtrochanteric level of paired dog femurs. On one side, the femur was fixed with Kirschner wires and a wire cerclage as described previously. For the other femur, the osteotomy was fixed with the angled plate. We found that both types of fixation presented the same stability in flexion-compression tests. However, under torsion the tension band wire fixation was 30%-50% less stable than the plate fixation.

Entities:  

Mesh:

Year:  1997        PMID: 9177801     DOI: 10.1007/BF00390050

Source DB:  PubMed          Journal:  Arch Orthop Trauma Surg        ISSN: 0936-8051            Impact factor:   3.067


  3 in total

1.  The outcome of paediatric LCP hip plate use in children with and without neuromuscular disease.

Authors:  S U Islam; A Henry; T Khan; N Davis; M Zenios
Journal:  Musculoskelet Surg       Date:  2013-12-03

2.  Development and mechanical testing of a short intramedullary nail for fixation of femoral rotational osteotomy in cerebral palsy patients.

Authors:  Rodrigo G Pagnano; Rodrigo Okubo; Jose B Volpon
Journal:  Biomed Eng Online       Date:  2011-06-28       Impact factor: 2.819

3.  Transfixing Kirshner wires for fixation of intertrochanteric valgus osteotomies in management of pediatric coxa vara.

Authors:  Ahmed Shawkat Rizk
Journal:  J Orthop Traumatol       Date:  2017-07-12
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.