Literature DB >> 9166306

The reliability of performing a screening cystometrogram using a fetal monitoring device for the detection of detrusor instability.

S E Swift1.   

Abstract

OBJECTIVE: To determine if a cystometrogram performed using the intrauterine pressure channel of a fetal monitor can be used to screen for detrusor instability in women undergoing evaluation for urinary incontinence.
METHODS: Sixty-six women with primary complaints of urinary incontinence were randomized to have a cystometrogram performed at their initial visit with a multichannel electronic cystometer or with the intrauterine pressure channel of a fetal monitor. Subjects underwent a second cystometrogram 1-4 weeks later with the alternative technique. The results were analyzed with chi 2 analysis and correlation coefficients.
RESULTS: Twenty-two subjects had detrusor instability diagnosed by the multichannel electronic cystometer. The fetal monitor cystometrogram was 91% sensitive and 86% specific in detecting detrusor instability, and had a 77% positive predictive value and a 95% negative predictive value. The correlation coefficients between the two examinations for bladder volume at first sensation, maximum capacity, volume at first contraction, and intensity of uninhibited detrusor contraction were r = 0.51, r = 0.69, r = 0.87, and r = 0.79, respectively; all of these were statistically significant.
CONCLUSION: The intrauterine pressure channel of a fetal monitor can be used reliably to perform a cystometrogram to screen for detrusor instability in patients presenting with complaints of urinary incontinence.

Entities:  

Mesh:

Year:  1997        PMID: 9166306     DOI: 10.1016/s0029-7844(97)00098-7

Source DB:  PubMed          Journal:  Obstet Gynecol        ISSN: 0029-7844            Impact factor:   7.661


  2 in total

1.  A comparison of maximum cystometric bladder capacity with maximum environmental voided volumes.

Authors:  E Yoon; S Swift
Journal:  Int Urogynecol J Pelvic Floor Dysfunct       Date:  1998

Review 2.  Urodynamic studies for management of urinary incontinence in children and adults.

Authors:  Keiran David Clement; Marie Carmela M Lapitan; Muhammad Imran Omar; Cathryn M A Glazener
Journal:  Cochrane Database Syst Rev       Date:  2013-10-29
  2 in total