Semi-conductor detectors in output factor measurements.

M G Karlsson1, M Karlsson, R Sjögren, H Svensson.   

Abstract

BACKGROUND AND
PURPOSE: Output factors are generally measured with cylindrical ionization chambers. It was investigated if Si-diodes of p-type instead could be used. The advantage would be the small detector size and the robust construction of the detector.
MATERIALS AND METHODS: Two types of diodes were studied, one with a shielding layer of tungsten specially made to reduce the excess response for scattered photons and one standard diode without any extra shielding. The measurements were performed at accelerating potentials between 4 and 50 MV and beam sizes between 4 cm x 4 cm and 40 cm x 40 cm.
RESULTS: The results showed that both types of diodes are suitable for measurements of head scatter factors in mini-phantoms. However, the diodes were found inappropriate for measurement of output factors for large fields in extended water phantoms. For small fields (<10 cm x 10 cm) a small detector is advantageous and no errors due to the scatter contribution were seen.
CONCLUSIONS: An cylindrical ionization chamber is the best choice for output factor measurements in extended water phantoms for large field sizes while diodes are an alternative in small fields. There were negligible differences between the detectors in head scatter measurements in mini phantoms.

Entities:  

Mesh:

Substances:

Year:  1997        PMID: 9155081     DOI: 10.1016/s0167-8140(97)01917-8

Source DB:  PubMed          Journal:  Radiother Oncol        ISSN: 0167-8140            Impact factor:   6.280


  1 in total

1.  Evaluation of a new VMAT QA device, or the "X" and "O" array geometries.

Authors:  Vladimir Feygelman; Geoffrey Zhang; Craig Stevens; Benjamin E Nelms
Journal:  J Appl Clin Med Phys       Date:  2011-01-31       Impact factor: 2.102

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.