Effects of four different methods of sampling arterial blood and storage time on gas tensions and shunt calculation in the 100% oxygen test.

F W Smeenk1, J D Janssen, B J Arends, G A Harff, J A van den Bosch, J P Schönberger, P E Postmus.   

Abstract

At the present time, plastic syringes are most commonly used for collecting arterial blood. The oxygen tension of the arterial blood (Pa,O2) in these syringes may fall. We studied the effect of the type of syringe, metabolism, and storage time on the arterial oxygen pressures measured and on the pulmonary shunt calculated. In 10 patients, 2-3 h after aortacoronary bypass surgery, a 100% oxygen test was performed. Four arterial blood gas samples were withdrawn from each patient in random order, two in glass syringes and two in plastic syringes. One glass and one plastic syringe were stored at room temperature (RT), and the others were stored in ice-water (IW). Each sample was analysed as soon as possible, and repeated 15, 30, 60 and 120 min after sampling. The Pa,O2 measurement in blood in the glass syringe in IW measured as soon as possible after sampling was considered the "gold standard". Pulmonary shunt calculations were performed using the results of the various blood gas analyses. Compared with the "gold standard", all of the other methods showed significant deterioration in the Pa,O2 measurement. The effect due to diffusion was 0.05 kPa x min(-1), and that due to metabolism 0.11 kPa x min(-1). The Pa,O2 in the glass syringes stored in IW remained stable with time. The pulmonary shunt was significantly overestimated when the "gold standard" blood gas results were not used (range 0.8-9.9%). Glass (not plastic) syringes should be used in the 100% oxygen test. The syringe should be cooled immediately, even when the sample is analysed as soon as possible.

Entities:  

Mesh:

Substances:

Year:  1997        PMID: 9150334

Source DB:  PubMed          Journal:  Eur Respir J        ISSN: 0903-1936            Impact factor:   16.671


  3 in total

1.  Effect of Sample Storage Temperature and Time Delay on Blood Gases, Bicarbonate and pH in Human Arterial Blood Samples.

Authors:  Elham Mohammadhoseini; Enayat Safavi; Sepideh Seifi; Soroush Seifirad; Shahram Firoozbakhsh; Soheil Peiman
Journal:  Iran Red Crescent Med J       Date:  2015-03-20       Impact factor: 0.611

Review 2.  Preanalytical considerations in blood gas analysis.

Authors:  Geoffrey Baird
Journal:  Biochem Med (Zagreb)       Date:  2013       Impact factor: 2.313

3.  Stability of pH, Blood Gas Partial Pressure, Hemoglobin Oxygen Saturation Fraction, and Lactate Concentration.

Authors:  Ariadna Arbiol-Roca; Claudia Elizabeth Imperiali; Dolors Dot-Bach; José Valero-Politi; Macarena Dastis-Arias
Journal:  Ann Lab Med       Date:  2020-06-17       Impact factor: 3.464