Improvement of platelet storage conditions by using new polyolefin containers.

J de Wildt-Eggen1, J G Schrijver, H J Bouter-Valk, R Fijnheer, M Bins, H C van Prooijen.   

Abstract

BACKGROUND: Storage of pooled platelet concentrates (PCs) with yields above 3.0 x 10(11) platelets per unit in a 1-L PL-732 polyolefin container for 5 days often results in a drop in pH to below 6.0. Recently, new oxygen-permeable platelet containers (1-L PL-2410, 1-L and 1.5-L Compoflex) have been developed. The maximal platelet storage capacities of the new containers and the PL-732 were compared. STUDY DESIGN AND METHODS: Large platelet pools (n = 27) with platelet concentrations between 1.2 and 1.4 x 10(11) per L were made from 3 to 5 PCs prepared from buffy coats. The pools were divided in equal volumes among the PL-732 and the three new platelet containers. Platelet counts in the PCs ranged from 1.0 to 5.0 x 10(11) per unit. All PCs were stored on a flatbed shaker at 22 +/- 2 degrees C and evaluated on Days 1, 3, 5, and 7 by measuring platelet count, pH, pO2, pCO2, HCO3-, glucose, lactate, platelet swirling, and soluble p-selectin.
RESULTS: Day 7 storage of PCs (n = 6) with yields between 3.0 and 4.0 x 10(11) platelets in PL-732 showed mean +/- SD pH values of 5.93 +/- 0.05 and lactate values of 32.3 +/- 7.9 mmol per L; in 4 of these 6 PCs, pH was below 6.0. In contrast, storage of these PCs in 1-L PL-2410 and 1.5-L Compoflex containers and of 2 of these 6 PCs in 1-L Compoflex containers showed pH values above 6.8. Lactate values were 15.5 +/- 1.3, 15.3 +/- 1.8, and 19.5 +/- 4.7 mmol per L, respectively (p < 0.001 vs. PL-732). The platelet storage capacity of the new containers with platelet yields between 4.0 and 5.0 x 10(11) per unit (n = 6) was evaluated. Day 7 storage of these PCs in the 1.5-L Compoflex showed an average pH value of 6.74 +/- 0.20; in 2 of 6 PCs, pH was below 6.8. The average pH value in the PL-2410 was 6.38 +/- 0.31, and in all PCs, pH was below 6.8. Average lactate values were 17.8 +/- 5.7 and 25.8 +/- 5.6 mmol per L (p < 0.05), respectively. Soluble p-selectin values on Day 7 of storage increased approximately twofold in all PCs.
CONCLUSION: The new oxygen-permeable containers showed platelet quality comparable to that with the PL-732 and for longer storage periods and at higher platelet counts.

Entities:  

Mesh:

Substances:

Year:  1997        PMID: 9149770     DOI: 10.1046/j.1537-2995.1997.37597293876.x

Source DB:  PubMed          Journal:  Transfusion        ISSN: 0041-1132            Impact factor:   3.157


  3 in total

1.  10 Years Experience with Bacterial Screening of Platelet Concentrates in the Netherlands.

Authors:  Dirk de Korte
Journal:  Transfus Med Hemother       Date:  2011-07-08       Impact factor: 3.747

2.  Exploratory studies of extended storage of apheresis platelets in a platelet additive solution (PAS).

Authors:  Sherrill J Slichter; Jill Corson; Mary Kay Jones; Todd Christoffel; Esther Pellham; S Lawrence Bailey; Doug Bolgiano
Journal:  Blood       Date:  2013-11-20       Impact factor: 22.113

3.  Platelet-rich fibrin prepared from stored whole-blood samples.

Authors:  Kazushige Isobe; Masashi Suzuki; Taisuke Watanabe; Yutaka Kitamura; Taiji Suzuki; Hideo Kawabata; Masayuki Nakamura; Toshimitsu Okudera; Hajime Okudera; Kohya Uematsu; Koh Nakata; Takaaki Tanaka; Tomoyuki Kawase
Journal:  Int J Implant Dent       Date:  2017-03-01
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.