| Literature DB >> 9146495 |
S S Yoo1, C R Guttmann, J R Ives, L P Panych, R Kikinis, D L Schomer, F A Jolesz.
Abstract
A method to visualize surface EEG electrodes on conventional high-resolution magnetic resonance (MR) images is presented. Because conventional silver or gold electrodes generate an insignificant artifactual signal signature in MR images, a plastic capsule filled with gadolinium-doped water was designed to mold onto each electrode, thereby outlining the electrode surfaces as a negative imprint on MR images. MR images of the head with capsules affixed clearly demonstrated the location and shape of the electrodes. Selected cranial tissues were identified by using semi-automatic image segmentation to determine their three-dimensional spatial relationship to the electrodes. Such representations could be used as a starting point for more precise EEG source localization modeling applicable to individual patients.Entities:
Mesh:
Year: 1997 PMID: 9146495 DOI: 10.1016/s0013-4694(96)95088-9
Source DB: PubMed Journal: Electroencephalogr Clin Neurophysiol ISSN: 0013-4694