Literature DB >> 9146328

Endothelin-1 and neutrophil activation during heparin-coated cardiopulmonary bypass.

R Lundblad1, O Moen, E Fosse.   

Abstract

BACKGROUND: Heparin-coated circuits attenuate the systemic inflammatory response to cardiopulmonary bypass. The present study compares two different heparin coatings in terms of the release of endothelin-1 and neutrophil glycoproteins.
METHODS: Forty low-risk patients undergoing coronary artery bypass grafting were investigated, having cardiopulmonary bypass with a Duraflo II heparin-coated circuit (n = 10), an identical but uncoated circuit (n = 10), a Carmeda BioActive Surface heparin-coated circuit (n = 10), or an identical but uncoated circuit (n = 10). A standard systemic heparin dosage was used in all patients. Endothelin-1 and the neutrophil glycoproteins lactoferrin and myeloperoxidase were quantified throughout the operation and 3 hours postoperatively.
RESULTS: Enhanced plasma levels of endothelin-1, lactoferrin, and myeloperoxidase were observed during and after uncoated cardiopulmonary bypass, but this was not associated with clinical side effects. Compared with the respective uncoated controls, Duraflo II attenuated only the lactoferrin levels, whereas Carmeda BioActive Surface was associated with lower levels of both endothelin-1, lactoferrin, and myeloperoxidase. Of the two heparin coatings, Carmeda BioActive Surface proved more effective than Duraflo II in attenuating the levels of these substances.
CONCLUSIONS: The plasma levels of endothelin-1, lactoferrin, and myeloperoxidase increase during cardiopulmonary bypass in coronary artery bypass grafting, but this has no clinical side effects in low-risk patients. The increase is attenuated using heparin-coated extracorporeal circuits, and then more effectively by Carmeda BioActive Surface than by Duraflo II.

Entities:  

Mesh:

Substances:

Year:  1997        PMID: 9146328     DOI: 10.1016/s0003-4975(97)00088-x

Source DB:  PubMed          Journal:  Ann Thorac Surg        ISSN: 0003-4975            Impact factor:   4.330


  4 in total

1.  Off-pump myocardial revascularization attenuates endothelin-1 expression in systemic, pulmonary, and coronary circulation.

Authors:  Daniel Unić; Davor Barić; Kristina Brkić; Mislav Planinc; Dubravka Jonjić; Igor Rudež; Željko Sutlić
Journal:  Wien Klin Wochenschr       Date:  2014-11-15       Impact factor: 1.704

2.  Treatment of coronary heart disease with minimally invasive surgery.

Authors:  M A Wait
Journal:  Proc (Bayl Univ Med Cent)       Date:  2000-04

3.  Cardiopulmonary Complications Following Cardiac Surgery.

Authors:  Renée L. Quarterman; Arthur Wallace; Mark B. Ratcliffe
Journal:  Curr Treat Options Cardiovasc Med       Date:  2001-04

Review 4.  Overview of lung transplantation.

Authors:  Marlyn S Woo
Journal:  Clin Rev Allergy Immunol       Date:  2008-12       Impact factor: 8.667

  4 in total