Unfavorable mechanical effects of heat and moisture exchangers in ventilated patients.

G A Iotti1, M C Olivei, A Palo, C Galbusera, R Veronesi, A Comelli, J X Brunner, A Braschi.   

Abstract

OBJECTIVE: To investigate the mechanical effects of artificial noses.
SETTING: A general intensive care unit of a university hospital. PATIENTS: 10 patients in pressure support ventilation for acute respiratory failure.
INTERVENTIONS: The following three conditions were randomly tested on each patient: the use of a heated humidifier (control condition), the use of a heat and moisture exchanger without filtering function (HME), and the use of a combined heat and moisture exchanger and mechanical filter (HMEF). The pressure support level was automatically adapted by means of a closed-loop control in order to obtain constancy, throughout the study, of patient inspiratory effort as evaluated from airway occlusion pressure at 0.1 s (P0.1). Patient's ventilatory pattern, P0.1, work of breathing, and blood gases were recorded.
MEASUREMENTS AND MAIN RESULTS: The artificial noses increased different components of the inspiratory load: inspiratory resistance, ventilation requirements (due to increased dead space ventilation), and dynamic intrinsic positive end-expiratory pressure (PEEP). The additional load imposed by the artificial noses was entirely undertaken by the ventilator, being the closed-loop control of P0.1 effective to maintain constancy of patient inspiratory work by means of adequate increases in pressure support level.
CONCLUSIONS: The artificial noses cause unfavorable mechanical effects by increasing inspiratory resistance, ventilation requirements, and dynamic intrinsic PEEP. Clinicians should consider these effects when setting mechanical ventilation and when assessing patients' ability to breathe spontaneously.

Entities:  

Mesh:

Year:  1997        PMID: 9142578     DOI: 10.1007/s001340050347

Source DB:  PubMed          Journal:  Intensive Care Med        ISSN: 0342-4642            Impact factor:   17.440


  13 in total

1.  Comparison of the effects of two humidifier systems on endotracheal tube resistance.

Authors:  Indalecio Morán; Belén Cabello; Elizabeth Manero; Jordi Mancebo
Journal:  Intensive Care Med       Date:  2011-09-27       Impact factor: 17.440

2.  The effect of a pediatric heat and moisture exchanger on dead space in healthy pediatric anesthesia.

Authors:  Min A Kwon
Journal:  Korean J Anesthesiol       Date:  2012-05-24

Review 3.  Weaning from mechanical ventilation.

Authors:  I Alía; A Esteban
Journal:  Crit Care       Date:  2000-02-18       Impact factor: 9.097

4.  Mechanical effects of heat-moisture exchangers in ventilated patients.

Authors:  G A Iotti; M C Olivei; A Braschi
Journal:  Crit Care       Date:  1999-09-23       Impact factor: 9.097

5.  Evaluation of an active humidification system for inspired gas.

Authors:  Nicolás G Roux; Gustavo A Plotnikow; Darío S Villalba; Emiliano Gogniat; Vivivana Feld; Noelia Ribero Vairo; Marisa Sartore; Mauro Bosso; José L Scapellato; Dante Intile; Fernando Planells; Diego Noval; Pablo Buñirigo; Ricardo Jofré; Ernesto Díaz Nielsen
Journal:  Clin Exp Otorhinolaryngol       Date:  2015-02-03       Impact factor: 3.372

6.  Influence of heat and moisture exchanger use on measurements performed with manovacuometer and respirometer in healthy adults.

Authors:  Jeanette Janaina Jaber Lucato; Thiago Marraccini Nogueira da Cunha; Sara Solange Oliveira Costa Rocha; Fernanda Maria Palmieri de Carvalho; Daniele Cristina Botega; Jamili Anbar Torquato; Ana Cristina Gimenes; Renato Fraga Righetti
Journal:  Multidiscip Respir Med       Date:  2015-12-19

7.  Evaluating humidity recovery efficiency of currently available heat and moisture exchangers: a respiratory system model study.

Authors:  Jeanette Janaina Jaber Lucato; Alexander Bernard Adams; Rogério Souza; Jamili Anbar Torquato; Carlos Roberto Ribeiro Carvalho; John J Marini
Journal:  Clinics (Sao Paulo)       Date:  2009       Impact factor: 2.365

Review 8.  Review of ventilatory techniques to optimize mechanical ventilation in acute exacerbation of chronic obstructive pulmonary disease.

Authors:  Raghu M Reddy; Kalpalatha K Guntupalli
Journal:  Int J Chron Obstruct Pulmon Dis       Date:  2007

Review 9.  Humidification during mechanical ventilation in the adult patient.

Authors:  Haitham S Al Ashry; Ariel M Modrykamien
Journal:  Biomed Res Int       Date:  2014-06-25       Impact factor: 3.411

10.  Ventilatory changes during the use of heat and moisture exchangers in patients submitted to mechanical ventilation with support pressure and adjustments in ventilation parameters to compensate for these possible changes: a self-controlled intervention study in humans.

Authors:  Jeanette Janaina Jaber Lucato; Thiago Marraccini Nogueira da Cunha; Aline Mela Dos Reis; Patricia Salerno de Almeida Picanço; Renata Cléia Claudino Barbosa; Joyce Liberali; Renato Fraga Righetti
Journal:  Rev Bras Ter Intensiva       Date:  2017 Apr-Jun
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.