| Literature DB >> 9129596 |
A Gil1, A González, R Dal-Ré, V Dominguez, P Astasio, L Aguilar.
Abstract
Diluted dried blood drops on filter paper were compared with serum samples as a specimen source for qualitative anti-HAV antibody determination by ELISA. A total of 298 serum samples and dried blood drops were collected from a population of healthy adolescents (15.3 +/- 1.2 years old). The prevalence of anti-HAV antibody obtained by testing serum samples was 7.7% (95% CI:4.8 10.1). Compared with serum sampling the sensitivity and specificity of diluted dried blood drops were 91.3 and 99.3%. The positive and negative predictive values were 91.3 and 99.3%, respectively, and the likelihood ratios of positive and negative results were 91 and 0.09. It is proposed that this test represents a reliable procedure for anti-HAV antibody testing.Entities:
Mesh:
Substances:
Year: 1997 PMID: 9129596 PMCID: PMC2808775 DOI: 10.1017/s0950268896007297
Source DB: PubMed Journal: Epidemiol Infect ISSN: 0950-2688 Impact factor: 2.451