Literature DB >> 9128128

Malfunctioning Starr-Edwards mitral valve 21 years after installation.

K Sakata1, S Ishikawa, A Ohtaki, Y Otani, M Suzuki, O Kawashima, Y Morishita.   

Abstract

Two cases of malfunctioning Starr-Edwards cloth-covered mitral valve prostheses requiring reoperation are presented. Both cases underwent successful surgical repair 21 years after the valve replacement. The causes were a disturbance of the poppet during the opening movement due to excessive tissue ingrowth and a paravalvular leak associated with a tear of the valve seat. Replacement of the Starr-Edwards valve prosthesis more than 20 years after the initial installation has not been reported.

Entities:  

Mesh:

Year:  1997        PMID: 9128128

Source DB:  PubMed          Journal:  J Cardiovasc Surg (Torino)        ISSN: 0021-9509            Impact factor:   1.888


  1 in total

1.  Reoperation for a Starr-Edwards aortic valve implanted 28 years earlier.

Authors:  Katsuhiko Yoshida; Genyo Fujii; Shuichi Suzuki; Hideki Ishida; Takeru Shimomura; Akio Matsuura
Journal:  Jpn J Thorac Cardiovasc Surg       Date:  2002-08
  1 in total