Literature DB >> 9122416

Iliac arteries: plain and heparin-coated Dacron-covered stent-grafts compared with noncovered metal stents--an experimental study.

K Schürmann1, D Vorwerk, R Uppenkamp, B Klosterhalfen, A Bücker, R W Günther.   

Abstract

PURPOSE: To compare lumen patency, neointimal thickening, and local inflammatory reaction with noncovered metal stents and plain and heparin-coated Dacron-covered stent-grafts.
MATERIALS AND METHODS: In each of 16 sheep, four prostheses were placed in the iliac arteries: a plain and a heparin-coated Dacron-covered Cragg stent-graft, a noncovered Cragg stent, and a noncovered Memotherm nitinol stent. Two prostheses were inserted in the iliac artery on each side. Patency was monitored with angiography and intravascular ultrasound (US). Eight animals were killed at 1-month follow-up, and the other eight at 6 month follow-up. Neointimal thickening and inflammatory foreign-body reaction to the prostheses were determined macroscopically and with light microscopy.
RESULTS: More than 50% stenosis developed in 10 heparin-coated and three plain Dacron-covered stent-grafts but not in the noncovered stents. Two heparin-coated and one plain Dacron-covered stent-graft occluded completely. Neointimal thickening and inflammatory reaction were severe with the heparin-coated Dacron-covered stent-grafts, were less with the plain Dacron-covered stent-grafts, and were least with the noncovered stents.
CONCLUSION: Neointimal thickening and inflammatory reaction significantly increased with covered compared with noncovered prostheses in the iliac arteries in sheep and were worst with heparin-coated Dacron-covered stent-grafts. If possible, use of noncovered stents should thus be preferred to use of Dacron-covered stent-grafts.

Entities:  

Mesh:

Substances:

Year:  1997        PMID: 9122416     DOI: 10.1148/radiology.203.1.9122416

Source DB:  PubMed          Journal:  Radiology        ISSN: 0033-8419            Impact factor:   11.105


  2 in total

1.  Development of in vivo tissue-engineered autologous tissue-covered stents (biocovered stents).

Authors:  Yasuhide Nakayama; Yue-Min Zhou; Hatsue Ishibashi-Ueda
Journal:  J Artif Organs       Date:  2007-09-20       Impact factor: 1.731

2.  DISCOVER: Dutch Iliac Stent trial: COVERed balloon-expandable versus uncovered balloon-expandable stents in the common iliac artery: study protocol for a randomized controlled trial.

Authors:  Joost A Bekken; Jan Albert Vos; Ruud A Aarts; Jean-Paul P M de Vries; Bram Fioole
Journal:  Trials       Date:  2012-11-19       Impact factor: 2.279

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.